This site uses cookies to provide an optimized shopping experience. By using this site, you agree the use of cookies within our privacy policy.
Fluke Calibration 9190A-E-P-156
Ultra-Cool Field Dry-Block Metrology Temperature Calibrator with Model 9190-INSE ("E" Insert) and Process Electronics
9190AEP156
9190A-E-P-156
4179006
Fluke Calibration 9190A-E-P-156
Ultra-Cool Field Dry-Block Metrology Temperature Calibrator with Model 9190-INSE ("E" Insert) and Process Electronics
9190AEP156
9190A-E-P-156
4179006
9190A
- Temperature range: -95 °C to 140 °C
- Built-in reference thermometer readout accuracy: ± 0.05 °C full range
- Display accuracy: ± 0.2 °C full range
- Stability: ± 0.015 °C full range
- Cooling time: 23 °C to -90 °C: 80 minutes; 23 °C to -95 °C: 90 minutes; 140 °C to 23 °C: 60 minutes
- Portability: Weighs 16 kg (35 lb)
- Built-in front and back handles for easy two-handed carry
- Conforms to EURAMET cg-13 guidance on measurement practices for temperature block calibrators
- All controls and inputs on front panel
- Stability indicator shows whether block temperature is stable and within limits defined by the user PC interface
- USB and RS-232 support remote operation from computer
- Active heater zone control compensates for differential temperatures between zones
- Minimizes axial temperature gradients
- Define up to eight set-point temperatures; automatically ramp to and "soak" at each
- Specify soak time, number of cycles, and direction
- Fast time to temperature and small footprint
- Compact, efficient free piston Stirling cooler technology (FPSC)
9190-INSE
- Six metric/imperial miscellaneous holes
- Hole quantities and sizes: 1 Hole, 3 mm; 1 Hole, 4 mm; 1 Hole, 6 mm; 1 Hole, 8 mm; 1 Hole, 10 mm; 1 Hole, 1/4 in.
- Removable inserts sized for probes with named diameter
- Use additional inserts to broaden range of hole sizes available
- For use with Model 9190A
Process Electronics
- Reference sensor control:
- Set-point temperature control gets passed from internal sensor to external reference PRT
- 9190A controls well temperature based on its measurement and position inside well
- Helps minimize effects of axial gradient when reference PRT is aligned with short sensors
- 4-20 mA connectors:
- Power 4-20 mA transmitter from drywell
- Supplies 24 V dc loop power
- 4-wire PRT/RTD input:
- Reads 4-wire, 3-wire, or 2-wire PRT/RTD
- Accuracy ± 0.02 °C
- 4-20 mA circuit fuses
- Thermocouple input:
- Accepts common thermocouples with mini-jack termination
- Types: J, K, T, E, R, S, M, L, U, N, C
- Reference thermometer input:
- Smart connector input accepts -A terminated reference PRTs
- Accuracy ± 0.01 °C at -95 °C