ACR Systems SRP-010-128K

SmartReader Plus 10 4-Channel (Temperature and RTD Temperature) Data Logger with 128 KB Memory

SRP010128K
SRP-010-128K
01-0132

ACR Systems SRP-010-128K

SmartReader Plus 10 4-Channel (Temperature and RTD Temperature) Data Logger with 128 KB Memory

SRP010128K
SRP-010-128K
01-0132

The ACR Systems SRP-010-128K SmartReader Plus 10 is a 4-Channel (Temperature, RTD Temperature) Data Logger that precisely measures temperature over a wider range than a thermistor without sacrificing accuracy or resolution. Setup and download information with the IC-102 interface cable. With built-in menus, logger setup is fast and easy.

  • Operate with TrendReader 2 Software
  • Memory Size: 128 KB
  • 12-bit resolution
  • Connect up to 10 data loggers
$1,262.00
This item ships FREE
  • Note: SmartReader Data Loggers require TrendReader 2 Software to operate
  • Note: Probes, transformers and transducers not included
  • Number of Channels: Four (4)
    • One (1) channel for ambient temperature
    • Three (3) channels for temperature (with RTD probes)
  • Monitor and record temperature in four different locations
  • Wide temperature range
  • Extreme precision and accuracy
  • Memory Size: 128 KB
  • 12-bit resolution
  • 10-year battery life
  • 3-year warranty
  • Fast sample rates
  • Remote access via modem or Ethernet
  • Network capable (connect up to 10 data loggers)
  • Alarm dial-out
  • User configurable
  • Case Material: Noryl plastic
  • External Connector: Removable screw-type terminal strip with Common (-) connection
  • Mounting: Magnetic backing or locking eyelet
  • Operating Limits: -40 to 70°C (-40 to 158°F) and 0 to 95% relative humidity (non-condensing)
  • Clock Accuracy: +/- 2 seconds per day
  • Battery: 3.6 volt Lithium, 1 Amp-Hour
  • Battery Life: 10 years under normal use (factory replaceable)
  • Power Consumption: 5 to 10 microamps (continuous)
  • Sampling Methods:
    • Continuous (First-in, First-out - not available with sample rates faster than 8 seconds)
    • Stop when full (Fill-then-stop)
  • Sampling Rates: User selectable rates from 25 per second (0.04 seconds) to once every 8 hrs
  • Resolution: 12-bit (1 part in 4,096)

Internal Temperature Sensor

  • Type: Negative Temperature Coefficient Thermistor 10,000 ohms @ 25°C (77°F)
  • Range: -40 to 70°C (-40 to 158°F)
  • Accuracy: +/- 0.2°C over the range of 0 to 70°C (+/- 0.3°F over range of 32 to 158°F)
  • Resolution: 0.03°C at 25°C (0.05°F at 77°F); better than 0.07°C (0.12°F) between -25 and 70°C (-13 and 158°F); better than 0.13°C (0.23°F) between -40 and -25°C (-40 and -13°F)
  • Internal Temperature Sensor
  • Type: Negative Temperature Coefficient Thermistor 10,000 ohms @ 25°C (77°F)
  • Range: -40 to 70°C (-40 to 158°F)
  • Accuracy: +/- 0.2°C over the range of 0 to 70°C (+/- 0.3°F over the range of 32 to 158°F)
  • Resolution: 0.03°C at 25°C (0.05°F at 77°F); better than 0.07°C (0.12°F) between -25 and 70°C (-13 and 158°F); better than 0.13°C (0.23°F) between -40 and -25°C (-40 and -13°F)

RTD Type

  • Any 100 or 1000 Ohm Platinum RTD with Temperature Coefficient of Resistance (TCR) of 0.00385
  • Accuracy (100 and 1000 Ohm): +/- 0.3°C (+/- 0.54°F) + resolution
  • Resolution (100 and 1000 Ohm): +/- 0.14°C (+/- 0.25°F)
  • Temperature Range (100 and 1000 Ohm): -230 to 300°C (-382 to 572°F)
  • Wiring Configurations: 2, 3, or 4 wire RTDs

Applications

  • Food process verification
  • Pharmaceutical storage monitoring
  • Transportation of temperature-sensitive goods