Models |
Range |
|
Measuring Arm |
0 to 11, 111Ω adjustable in 0.1Ω steps,
(consists of five decades; 1,000Ω x 10 + 100Ω x 10Ω x 10
+ 1Ω x 10 + 0.1Ω x 10) |
|
Multiplier |
x 100mΩ, x 1Ω, x 10Ω, x100Ω, x1kΩ, x
10kΩ, x100kΩ, x1MΩ, and x10MΩ |
|
Accuracy |
(at 23±2°, less then 75%) |
|
x 100mΩ range.. ±0.5mΩ, x 1Ω range...
(±0.02% of setting +0.5mΩ) |
|
x 10Ω range, x 100kΩ & 1MΩ
ranges... ±0.02% of setting |
|
x100Ω x 1kΩ & x 10kΩ ranges... ±0.01% of
setting |
|
x10MΩ range... ±0.05% of setting |
|
Maximum Allowable Input |
0.2W continuously for ratio arm, 0.2W
continuously per element of Measuring arms |
|
Galvanometer (built-in) |
Max sensitivity... approx. 10µV/div.
(provided with a sensitivity adjuster) |
|
Galvanometer Battery |
Single 9V battery (JIS 6F22 or
equivalent) |
|
Galvanometer
Battery life |
Approx. 300 Hours |
|
Insulation Resistance |
More then 1000Ω at 250V DC at ambient
Humidity of less then 75% between electric circuit and
case |
|
Dielectric Strength |
500V AC for one minute between electric
circuit and case |
|
Bridge power source (built-in) |
Power source range |
1.5/3/6/15V selectable, |
|
Power source battery |
four 1.5V batteries (JIS R20P, ANSI
D, Mono 1.5V or equivalent) and a single 9V battery (JIS
6F22, ANSI 6F22, Energieblock 9V or equivalent) |
|
Case |
Grey metal Case, with plastic feet and
carrying handles |