Parameter |
Reference |
Accuracy |
|
DC Current |
400/2000A |
±1.5+2 |
|
AC Current |
400/2000A(150~1700A) |
1.5+3 (50/60Hz) |
|
3.0+4 (30~1kHz) |
|
2000A (1001~2000A) |
3.5+3 (50/60Hz) |
|
DC voltage |
40/400/1000V |
1.0+2 |
|
AC Voltage |
40/400/750V |
1.5+3 (50/60Hz) |
|
2.0+4 (30~1kHz) |
|
Crest Factor |
|
≤3 |
|
Resistance |
400/4000Ω |
1.5+2, Beeps at below 50+35Ω (Continuity
check) |
|
Frequency |
10-3999 |
1.5+5 |
|
DC Output |
DC400A(0~400mV) |
±1.5%rdg ±3mV |
|
DC2000A(15~200mV) |
±1.5%rdg ±3mV |
AC400A(0~400mV)
AC2000A(15~170mV/150~1700A) |
±1.5%±3mV (50/60Hz) |
|
±3.0%rdg ±3mV (40~1kHz) |
|
AC2000A(170.1~200mV/ 1701~2000A) |
±3.5%±3mV (50/60Hz) |
|
|
|
Parameter |
Specification |
|
Method of Detection |
True RMS |
|
Diameter of Measurable conductor |
Ø55MM at maximum |
|
Display |
LCD (Digital Display: 3999 Counts) |
Response Time
(Approx.) |
1 Seconds (on DC currency/voltage Range)
2 seconds (on AC current/Voltage Range, Resistance
Range) |
|
Peak Hold |
On AC Current Range |
|
Data Hold |
On All range |
|
Ambient Temperature and Humidity |
0-40˚C, 85% RH or less (no condensation) |
|
Influence of external Magnetic Field |
4A or less at 400 A/m |
|
Influence of Conductor Position |
±(1.5%rdg + 3dgt) or Less |
|
Circuit Voltage |
1000Vrms or less |
|
Withstanding Voltage |
5.32kV AC for one minute |
|
Power Supply |
6F22(006P)9v x1; Or 6RL61 x1 |
|
Battery Life |
Approx. 15 hours (continuous) |
|
Consumed Current |
Approx. 15mA |
|
Sleep Function |
Automatically powered down in about 10
minutes after last switch operation |
|
Dimensions |
Approx. 105(W)x247(H)x49(D)mm |
|
Weight |
Approx. 540g |