Dielectric Analyzer Specifications

AC DIELECTRIC MODE

AC Voltage Output
Range  Current Accuracy
100 to 500 VAC 20 mA     0.7% of output + 7.5V
500 to 5000 VAC  35 mA 0.7% of output + 30V
Output Frequency 50Hz to 100Hz with 1Hz resolution independent of line power frequency
Output Waveform Digitally synthesized, low distortion sinewave
AC Current Sensing (AC Leakage)
Range    Resolution   Accuracy
13.00uA 10nA 0.5% of reading + 1uA
1.300mA 1uA 0.5% of reading + 10uA
40.0mA 100uA 0.5% of reading + 500uA
Current Sensing Resistive current (in-phase), reactive current (out-of-phase) or total current
Voltage Coefficient 100pA/V
DC DIELECTRIC MODE
DC Voltage Output
Range  Current   Accuracy
100 to 500 VDC 20mA  0.3% of output + 6V
500 to 5,000 VDC 35mA 0.3% of output + 25V
5,000 to 7,000 VDC* 5mA 0.3% of output + 75V
Requires option *DC-7  
DC Voltage Output with option DC-10
500 to 10,000 VDC 5mA 1% of output + 50V
DC Current Sensing (DC Leakage)
Range Resolution Accuracy
1.800uA    1nA  0.4% of reading + 6nA
18.00uA 10nA 0.3% of reading + 30nA
1.800mA 1uA 0.3% of reading + 30uA
40.0mA 10uA 0.3% of reading + 300uA
Voltage Coefficient 15pA/V
HIGH RESISTANCE MODE (Insulation Resistance)
Resistance Range Minimum Resistance Test Voltage
2.6 GΩ 100 KΩ   100 VDC
13 GΩ   500 KΩ 500 VDC
26 GΩ 1 MΩ  1,000 VDC
65 GΩ  2.5 MΩ 2,500 VDC
Accuracy Combined uncertainty of DC Voltage Output + Current Measurement Accuracy  
LOW RESISTANCE MODE (Earth Continuity)
Full Scale 8.5 ohms maximum including lead resistance
Resolution 1 milli-ohms
Test Current 50 mA
Lead Compensation Automatic offset circuit "nulls" out two wire lead resistance
Accuracy 0.3% of reading + 30mΩ (up to 5Ω)
General Specifications
Arc Detection Sensitivity User selectable 3 microseconds to 30 microseconds
Display 40 character alphanumeric vacuum fluorescent
High Voltage Indicator Lightning bolt illuminated by independent output voltage monitor.
Audible alarm sounds prior to voltage output.
Voltage Resolution 1 volt AC or DC
Dwell Time Accuracy +2 Seconds/minute, -500mS/minute
Slew Rate 10 Volts/second to 1999 volts/second, user selectable
Limits Test automatically aborts when digitally stored minimum or maximum limit is exceeded.
Remote Interface Option "IF-4" GPIB - IEE-488.2 Subset compatible
RS232C - 1200, 2400, 4800, 9600 Baud rate selectable
Parallel - IBM/Centronics compatible
VICL - Vitrek auxiliary unit control loop
Accuracy Stated specifications are bipolar and apply for one year within "5EC from calibration temperature (20 to 25EC).  
Temperature Coefficient 0.1 times stated accuracy per EC beyond calibration temperature window
Calibration Traceable to the National Institute of Standards and Technology (NIST).
Automatic "Covers on" calibration, internal or external.
Calibration constants stored in non-volatile memory.
Warm up Time 30 minutes
Operating Temperature 0°C to 50°C
Humidity 90% RH, 0° to 40° C
Power 115/230 VAC + 10%, 350 VA Max.
Dimensions 89mm H x 432mm W x 457mm D (3.5" H x 17" W x 18" D
Weight 14Kg (34 lb.) Net / 18Kg (40 lb.) shipping

944i Product Page

Vitrek Catalog   |   Vitrek Price List

$20.00 Minimum on Total Order