| Alternating
Current Output |
| Range |
3 to 30
amperes RMS, programmable |
| Resolution |
1 amp |
| Output
Frequency |
Sinewave
synthesis, 50 Hz to 400Hz |
| Compliance
Voltage |
3 volts
RMS |
| Regulation |
± 0.25%
no load to full load |
| Resistance
Measurement |
| Resistance
Limit |
User
programmable max. and min. Test is automatically aborted when limit
is exceeded. |
| Resistance
Range |
1Ω
at 3 amps linearly decreasing to 100 mΩ
at 30A |
| Resistance
Resolution |
10mΩ |
| Measurement
Accuracy |
± 1.0% of reading; ± 0.5% of
range |
| Accuracy |
Specifications
apply for 1 year, +/- 5°C from the cal temp within 20°C to 30°C |
| Calibration |
Certificate
of calibration traceable to the National Institute of Standards and
Technology (NIST) provided |
| Test
Time |
Digital
count down timer, programmable 1 to 9999 seconds |
| Terminal
Configuration |
True
four-wire kelvin |
| General
Specifications |
| Remote
Interface |
VICL and
serial com standard. PLC optional. |
| Display |
40
character dot matrix illuminated liquid crystal display |
| Indicators |
L.E.D.
status indicators |
| Alarm |
Audible
and visual alarms provide pass/fail indication |
| Diagnostics |
Built-in
self diagnostics test ROM, RAM and output system |
| Safety |
CE mark
certified to EN61010 |
| Accessories |
K-2,
VIC-3, BBL-2 and operators manual supplied with unit |
| Warranty |
Extended
5 year parts and labor with annual factory calibration and maintenance.
Standard warranty one year. |
| Operating
Temperature |
0°C to
50°C |
| Humidity |
90% RH 0°C
to 40°C |
| Power |
115/230
VAC +/- 10% 250 VA max |
| Dimensions |
89mmH x
432 mmW x 457mmD (3.5"H x 17"W x 18"D) |
| Weight |
11Kg (24 lb.) Net/15Kg (34
lb.) shipping |