HP 8591A SPECIFICATIONS
GENERAL
Temperature 0° to +55°C operating; -40° to 75° C storage
EMI compatibility CISPR Pub. 11 and FTZ 526/527/79
Audible noise < 37.5 dBA pressure and < 5.0 Bels power (IS0DP7779)
Power requirements 86 to 127 or 195 to 250 Vrms, 47 to 66 Hz. 103 to 126 Vrms, 400 Hz ± 10%
Power consumption < 300 VA; < 100 watts
Frequency
Range 9 kHz to 1.8 GHz (50 Ω); 1 MHz to 1.8 GHz (75 Ω, opt. 001)
Reference
Aging ± 2 X 10 -6/year
Settability ± 1 X 10 -8
Precision frequency reference (option 004)
Aging ±2 X 10 -7/year
Settability ± 1 X 10 -8
FREQUENCY
Frequency readout accuracy (start, stop, center, frequency)
Span ± 10 MHz ± (freq readout X freq ref error + 3% of span + 20% of RBW + 100 Hz)
Span ± 10 MHz ± (freq readout X freq ref error + 3% of span + 20% of RBW + 100 Hz)
Marker count accuracy (signal to noise ratio ± 25 dB, RBW/span ± 0.01)
Frequency span < 10 MHz ± (marker freq X freq ref error + counter res + 100 Hz)
Frequency span > 10 MHz ± (marker freq X freq ref error + counter res + 1 kHz)
Counter resolution selectable from 10 Hz to 100 kHz
Frequency span
Range 0 Hz (zero span), 10 kHz to 1.8 GHz
Accuracy ± 2% of span, span ± 10 MHz, ± 3% of span, span > 10 MHz
Sweep time
Range 20 ms to 100 s, span = 0 Hz or > 10 kHz; 20 µs to 100 s, span = 0 Hz (option 101)
Accuracy ± 3%, 20 ms to 100 s; ± 2% 20 µs to <20 ms (opt 101)
Sweep trigger free run, single, line, video, external
Stability
Noise sidebands ± 90 dBc/Hz at > 10 kHz offset from CW signal (1 kHz RBW, 30 Hz VBW, sample detector); ± 100 dBc/Hz at > 30 kHz offset from CW signal (1 kHz RBW, 30 Hz VBW, sample detector)
Residual FM < 250 Hz p-p in 100 ms (1 kHz RBW, 1 kHz VBW)
System-related sidebands <-65 dBc at > 30 kHz offset from CW
AMPLITUDE
Amplitude range -115 to +30 dBm (50 ohm); -63 to +75 dBmV (75 ohm, opt 001)
MAXIMUM SAFE INPUT 50 ohm 75 ohm
Average cont power +30 dBm (1 watt) +75 dBmV (0.4 watts)
Peak pulse power +30 dBm (1 watt) +75 dBmV (0.4 watts)
dc 25 Vdc 100 Vdc
Gain compression, >10 MHz ± 0.5 dB, total power at input mixer = -10 dBm
Displayed average noise level 115 to 113 dBm
NOISE LEVEL
Spurious responses
Second harmonic distortion 5 MHz to 1.8 GHz, < -70 dBc for -45 dBm tone at input mixer
Third-order intermodulation distortion 5 MHz to 1.8 GHz, <-70 dBc for two -30 dBm tones at input mixer and > 50 kHz sep
Other input-related spurious < -65 dBc for 30 kHz offset from CW signal
Residual responses (input terminated and 0 dB attenuation)
150 kHz to 1 MHz < -90 dBm, 50 ohm
1 MHz to 1.8 GHz < -90 dBm, 50 ohm; < -38 dBmV, 75 ohm
Display range
Log scale 0 to -70 dB from ref lev is cal'd; 1 to 20 dB/div in 1 dB steps; 8 divisions displayed
Linear scale 8 divisions
Scale units dBmV, dBmicroV, volts, watts
Marker readout resolution 0.05 dB, log scale; 0.07% of reference level, linear scale
Fast sweep times for zero span (opt 101) 20 µs to 20 ms, 0.7% of ref level for linear scale
Reference level
Range -115 to +30 dBm (50 ohm), -63 to +75 dBmV (75 ohm)
Resolution 0.01 dB for log scale; 0.12% of ref level for linear scale
Accuracy, referred to -20 dBm ref level 0 dBm to -59.9 dBm, ± (0.5 dB + input atten acc @ 50 MHz); -60 dBm to -115 dBm, ± (1.25 dB +input atten acc @ 50 MHz)
Frequency response
Absolute ± 1.5 dB, referred to 300 MHz CAL OUT
Relative flatness ± 1.0 dB, referred to midpoint between highest and lowest response deviations
Calibrator output
Frequency 300 MHz ± (300 MHz X freq ref error)
Amplitude -20 dBm ± 0.4 dB (50 Ω); +28.75 dBmV ± 0.4 dB (75 Ω, opt 001)
Input attenuator
Range 0 to 60 dB in 10 dB steps
Accuracy at 50 MHz, 10 dB atten ± 0.5 dB, 0 to 50 dB; ± 0.75 dB, 60 dB
Resolution bandwidth 1 kHz to 3 MHz, ± 20%
Switching uncertainty, ref to 3 kHz bandwidth 3 kHz to 3 MHz RBW, ± 0.4 dB; 1 kHz, ± 0.5 dB
Video bandwidth range 30 Hz to 1 MHz
Log to linear switching ± 0.25 dB at reference level
Display scale fidelity
Log incremental accuracy ± 0.2 dB/2 dB, 0 to -70 dB from ref lev
Log maximum cumulative ± 0.75 dB, 0 to -60 dB from ref level; ± 1.0 dB, 0 to -70 dB from ref level
Linear accuracy ± 3% of reference level

 

HP 8593A SPECIFICATIONS
FREQUENCY
Frequency range 9 kHz to 22 GHz; 9 kHz to 26.5 GHz (option 026)
Frequency reference
Aging ± 2 X 10 -6/year
Settability ± 5 X 10-7
Temperature stability ± 5 X 10-6
Precision frequency reference (Opt 004)
Aging ± 1 X 10-7/year
Settability ± 5 X 10-8
Temperature stability ± 1 X 10-8
Frequency readout accuracy ± (frequency readout X frequency reference error + 3% of span + 20% of RBW + 100 Hz sweep time) for spans ± 10 MHz; ± (freq readout X freq ref error + 3% of span + 20% of RBW) for spans > 10 MHz
Marker count accuracy (signal-to-noise ratio)± 25 dB, RBW/span ±0.01 ± (marker freq X freq ref error + counter res + 100 Hz) spans ± 10 MHz; ± (marker freq X freq ref error + counter res + 1 kHz), spans > 10 MHz
Counter resolution Selectable from 10 Hz to 100 kHz
Frequency span
Range zero span, (10 X N) kHz to 19.25 GHz, (10 X N) kHz to 23.75 GHz (opt 026)
Accuracy ± 2% of span, span < 10 MHz; ± 3% of span, span > 10 MHz
Sweep time
Range 20 ms to 100 s, span = 0 Hz or > 10 kHz; 20 µs to 100 s, span = 0 Hz (opt 101)
Accuracy ± 3%, 20 ms to 100 s, ± 2%, 20 µs to 20 ms
Sweep trigger free run, single, line, video, external
Stability
Noise sidebands -95 dBc/Hz + 20 log N at > 30 kHz offset from CW signal
Comb generator 100 MHz fundamental freq; ± 0.007% freq accuracy
AMPLITUDE
Amplitude range -114 to +30 dBm
Maximum safe input level +30 dBm (1 watt, 7.1 Vrms), 0 Vdc
Gain compression ± 0.5 dB (total power at input mixer = -10 dBm)
Displayed average noise level 114 to < -92 dBm
Spurious responses
Second harmonic distortion < -70 dBc for -40 dBm tone at input mixer, 10 MHz to 2.9 GHz; < -100 dBc for -10 dBm tone power at input mixer or below displayed av noise lev), > 2.75 GHz
Third-order intermodulation distortion > 10 MHz < -70 dBc for two -30 dBm tones at input mixer and > 50 kHz separation
Other input-related spurious < -70 dBc for applied freq & 18 GHz; < -60 dBc for applied freq & 22 GHz
Display range
Log scale 0 to -70 dB from ref lev is calibrated; 1 to 20 dB/div in 1 dB steps;  divisions displayed
Linear scale 8 divisions
Scale units dBm, dBmB, dBmicroV, volts, watts
Reference level
Range -114 to +30 dBm
Resolution 0.01 dB for log scale; 0.12% of ref lev for linear
Accuracy (ref to -20 dBm ref level) ± (0.05 dB + input atten acc @ 50 MHz), 0 dBm to -59.9 dBm, ± (1.25 dB + input atten acc @ 50 MHz), -60 to -114 dBm
Frequency response (ref to 300 MHz CAL OUT, preselector peaked)
Absolute ± 2.0 to ± 3.0 dB
Relative flatness ± 1.5 to ± 2.0 dB
Calibrator output
Frequency 300 MHz ± 30 kHz
Amplitude -20 dBm ± 0.4 dB
Input attenuator
Range 0 to 70 dB in 10 dB steps
Accuracy at 50 MHz, ref to 10 dB atten + 0.5 dB, 0 to 60 dB; + 1.2 dB, 70 dB
Resolution bandwidth 1 kHz to 3 MHz, ± 20%
Switching uncertianty ± 0.4 dB, 3 kHz to 3 MHz RBW; ± 0.5 dB, 1 kHz
Video bandwidth range 30 Hz to 1 MHz
Log to linear switching + 0.25 dB at reference level
Display scale fidelity ± 0.2 dB/ 2 dB, 0 to -70 from ref lev, incremental; ± 0.75 dB, 0 to -60 dB from ref lev and ± 1.0 dB, 0 to -70 dB from ref lev, maximum cumulative.
 
8591A, 8593A, 8594A, 8595A Product Page
 

Used Equipment Catalog

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved