| FREQUENCY CHARACTERISTICS |
| Range |
15 Hz to 50 kHz |
| Display |
5 digit LED readout |
| Resolution |
1 Hz |
| Accuracy |
± 3 Hz., 0 to 55°C |
| Typical Stability |
± 10 Hz/hour after 1 hour and ±5 Hz/°C |
| Automatic frequency control (AFC) hold-in range |
± 800 Hz |
| AMPLITUDE CHARACTERISTICS |
| Instrument Range |
| Linear |
30 V to 100 nV full scale |
| Log |
+30 dBm or dBV to -150 dBm or dBV |
| Amplitude Accuracy |
| Frequency response, 15 Hz - 50 kHz |
Log: ± 0.4 dB
Linear: ± 4% |
| Dynamic range |
> 80 dB |
| Noise Sidebands |
greater than 70 dB below CW signal. 10 bandwidths away
from signal. |
| Spurious responses |
> 80 dB below input reference level |
| SWEEP CHARACTERISTICS |
| Scan Width |
50 Hz to 50 kHz, adjustable in a 1-2-5 sequence from 50
Hz to the full frequency range |
| Sweep Error Light |
This LED indicates a sweep that is too fast to capture
full response. When the light is on, response will be lower than it should be. |
| External Trigger |
A short to ground stops the normal sweep. Opening the
short then enables a sweep. |
| INPUT CHARACTERISTICS |
| Impedance |
1 MΩ, 30 pF |
| Maximum Input Level |
100 V rms, ± 100 V dc |
| OUTPUT CHARACTERISTICS |
| Tracking generator output
(also known
as BFO or tracking oscillator output) |
| Range |
0 tp >1 V rms into 600
Ω |
| Frequency Response |
± 3% 15 Hz to 50 kHz |
| X-Y Recorder Analog Outputs |
| Vertical |
0 to +5 V ±2.5% |
| Horizontal |
0 to +5 V ±2.5% |
| Impedance |
1 kΩ |
| Recommended Accessory |
HP 7090A Measurement Plotting System |
| Pen Lift |
contact closure to ground during sweep |
| Restored Output |
acts as a narrow band amplifier |
| GENERAL |
| Power Requirements |
100V, 120V, 220V, or 240V +5% -10%, 48 Hz to 440 Hz, 10
VA typical |
| Size |
203.2 mm H X 285.8 mm W X 412.8 mm D (8"
X
11 1/4" X 16 1/4") |
| Weight |
11.5 kg (23 lb). Opt 001: 13.5 kg (30 lb) |