| FREQUENCY CHARACTERISTICS |
| Range |
5 Hz to 50 kHz |
| Digital Frequency Display |
| Resolution |
± 1 Hz |
| Accuracy |
± 3.5 Hz, 0 to 55° C |
| Typical Stability |
± 10 Hz/hr after 1 hour; ± 5 Hz/°C |
| Bandwidths (accuracy ± 15%) |
1 Hz (25°C ± 5C), 3 Hz, 10 Hz, 100 Hz, 300
Hz |
| Shape Factor |
1 Hz (25°C ± 5C), 3 Hz, 10 Hz, 100 Hz: 10
300 Hz: 8 |
| Out of range blank: if
controls are set so portions of displayed signal lie below 0 Hz or above 50 kHz, the
baseline is displayed. |
| AMPLITUDE CHARACTERISTICS |
| Overall Instrument Range |
Linear: 20 V -100 nV full scale
Log: +30 dBm or dB V;
-150 dBm or dB V |
| Amplitude Accuracy |
Log |
Linear |
| Frequency response |
| 20 Hz - 20 kHz |
± .3 dB |
± 3% |
| 5 Hz - 50 kHz |
± .5 dB |
± 5% |
| Switching between bandwidths
(25°C) |
| 3 Hz - 300 Hz |
± .5 dB |
± 5% |
| 1 Hz - 300 Hz |
± 1 dB |
± 10% |
| Amplitude display |
± 2 dB |
± 2% |
| Input attenuator |
± .3 dB |
± 3% |
| Amplitude reference level:
(IF
attenuator) |
| Most sensitive range |
± 1 dB |
± 10% |
| All other ranges |
± 1 dB |
± 3% |
| Dynamic Range |
80 dB |
| IF feedthru |
input level > 10V, -60 dB; < 10V, -70
dB |
| Spurious responses |
> 80 dB below input reference level |
| Smoothing |
3 positions, rolloff is a function of
bandwidth |
| Overload indicator |
this LED indicator warns of possible input
amplifier overloading. Without this indication it would be possible to introduce spurious
responses without knowing it. |
| SWEEP CHARACTERISTICS |
| Scan Width |
50 Hz to 50 kHz |
| Log Sweep |
20 Hz to 43 kHz ± 20% after 3 sweeps |
| Sweep Times |
0.1 s to 2000 s |
| Rep |
repetitive sweeps over the specified band |
| Reset |
resets to the beginning of the sweep-used to
adjust start or center frequency |
| Manual |
in combination with the concentric knob,
manual sweep fully duplicates the span of the electronic sweep |
| Adaptive Sweep |
when in adaptive sweep below the threshold
level, scan speed is 20 to 25 times faster. Threshold is adjustable to cover 0-60% of
screen. Signals greater than about 6 dB above threshold are detected and swept slowly. |
| Sweep Error Light |
this LED indicates a sweep that is too fast
to capture full response. When the light is on, response can be > 5% lower than it
should. |
| Zero Scan |
to look at the time varying signal at the
center or start frequency within the bandwidth selected, the zero scan is used. |
| OUTPUT CHARACTERISTICS |
| Tracking Generator Output
(also known as BFO or tracking oscillator output) |
| Range |
0 to 1 V rms into 600 Ω |
| Frequency Response |
± 3%, 5 Hz to 50 kHz |
| Impedance |
600 Ω |
| Total Harmonic and Spurious Content |
40 dB below 1 volt signal level |
| X-Y Recorder Analog Outputs |
| Vertical |
0 to +5 V ± 2.5% |
| Horizontal |
0 to +5 V ± 2.5% |
| Impedance |
1 kΩ |
| Pen Lift |
contact closure to ground during sweep |
| Recommended Accessory |
HP 7090A Measurement Plotting System |
| Size |
203.2 mm H X 285.8 mm W X
412.8 mm D (8" X 11 1/4 X
16 1/4") |
| Weight |
net, 12.25 kg (27 lb); HP 3580A Opt 001: net,
15.88 kg (35 lb) |
| Temperature Range |
0°C to 55°C |
| Power |
100V, 120V, 220V, or 240V +5% -10%; 48 to 440
Hz; 35 VA max |
| Opt 001 battery |
5 hours from full charge. 14 hours to fully
recharge. The internal battery is protected from deep discharge by an automatic turn off.
Useful life of batteries is over 100 cycles. |