Calibrating the Agilent 8902A Option 050 Increased Power Measurement
Accuracy, requires performance that is only available from a primary
standard. This entails using a precise piston attenuator to
calibrate the 8902A Option 050 to its specified power measurement
accuracy. For those customers that can not tolerate the turnaround
time associated with returning the instrument to the factory for
calibration or the expense of purchasing a primary standard, Agilent
Technologies is offering the 11812A Verification Kit. The 11812A
consists of a step attenuator, two 10 dB pads semi-permanently
attached, a cable, and a carrying case. The 11812 is calibrated
against the piston attenuator and is specified to verify the
performance of the 8902A Option 050 Tuned RF Level function to
±(0.015 d + 0.010 dB/ 10 dB step). |