HP 3466A is a 4 /2 digit autoranging multimeter offering six
functions ACV, DCV, ACI, DCI, Ohms, and Diode Test. ACV and
ACI measurements are true rms with selectable ac or dc coupling.
Available with rechargeable batteries or ac power only, it offers 1
µvolt and 1 mOhm sensitivity with zero adjustment on the lowest
ranges to compensate for external offsets. |

HP / Agilent 3466A Shown
|
|
Specifications |
|
VOLTAGE |
|
DCV |
| Range |
Maximum Display |
Accuracy: 1 year. 15 to 30 deg.C
± (% of reading + # of counts) |
| 20 mV |
± 19.999 |
0.05 + 3 |
| 200 mV |
± 199.99 |
0.04 + 2 |
| 2 V |
± 1.9999 |
0.03 + 1 |
| 20 V |
± 19.999 |
0.03 + 1 |
| 200 V |
± 199.99 |
0.035 + 1 |
| 1200 V |
± 1199.9 |
0.035 + 1, < 700 V input |
| 1200 V |
± 1199.9 |
0.055 + 1, > 700 V input |
| AC V (true-rms responding, true-rms calibrated) |
|
Frequency Range
|
Accuracy: 1 year. 15
to 30 deg.C
± (% of reading + # of counts)
|
| 20 to 30 Hz |
2 + 50
|
| 30 to 50 Hz |
1 + 30
|
| 50 Hz to 10 kHz |
0.3 + 20
|
| 10 to 20 kHz |
1 + 40
|
| 20 to 100 kHz |
2 + 150
|
| Maximum Input |
(ac + dc) V: ± 1200 Vdc, 1700 V (dc + peak
ac); acV: ± 600 V dc, 1700 V (peak ac + dc); 107 volt-Hz max. |
| Crest Factor |
4:1 at full scale |
| RESISTANCE |
20 Ohm to 20 MOhm in 7 ranges |
| Ranges |
250 V or 350 V (dc + peak ac) |
| Input Protection |
|
Range |
Accuracy: 1 year. 15
to 30 deg.C
± (% of reading + # counts) |
| 20 to 200 Ohm |
0.08 + 2 |
| 200 Ohm to 200 kOhm |
0.03 + 1 |
| 2000 kOhm |
0.04 + 1 |
| 20 MOhm |
0.15 + 1 |
|
CURRENT |
|
Maximum Input |
Current: 2 A (fused protected).
Voltage: 250
V |
| DCI |
| Range |
Accuracy: 1 year. 15
to 30 deg.C
± (% reading + # of counts) |
| 200 µA, 2 mA, 20 mA |
0.07 + 2 |
| 200 mA |
0.15 + 2 |
| 2000 mA |
0.5 + 2 |
| ACI |
| Range |
Frequency |
Accuracy: 1 year. 15 to 30 deg.C
± (% of reading + # of counts) |
| 200 µA to 200 mA |
20 to 30 Hz |
2.0 + 50 |
| 30 Hz to 10 kHz |
0.9 + 35
2.0 + 50 |
| 2000 mA |
20 to 30 Hz |
| 30 Hz to 10 kHz |
1.2 + 20 |
|