Vertical System Characteristics
  DPO4032
MSO4032
DPO4034
MSO4034
DPO4054
MSO4054
DPO4104
MSO4104
Input Channels 2 4 4 4
Analog Bandwidth
(-3 dB) 5 mV/div - 1 V/div
350 MHz 350 MHz 500 MHz 1 GHz
Calculated Rise Time
5 mV/div (typical)
1 ns 1 ns 700 ps 350 ps
Hardware Bandwidth Limits 20 MHz or 250 MHz
Input Coupling AC, DC, GND
Input Impedance 1 MΩ ±1%, 50 Ω ±1%
Input Sensitivity, 1 MΩ 1 mV/div to 10 V/div
Input Sensitivity, 50 Ω 1 mV/div to 1 V/div
Vertical Resolution 8 bits
Max Input Voltage, 1 MΩ 250 VRMS with peaks ≤±400 V
Max Input Voltage, 50 Ω 5 VRMS with peaks ≤±20 V
DC Gain Accuracy ±1.5% with offset set to 0 V
Offset Range(applies to all 4000 series models) 1MW 50W
1 mV/div – 50 mV/div  ±1 V
50.5 mV/div – 99.5 mV/div ±0.5 V
100 mV/div – 500 mV/div  ±10 V
505 mV/div – 995 mV/div ±5 V
1 V/div – 5 V/div ±100 V
5.05 V/div – 10 V/div ±50 V N/A
Channel-to-channel Isolation
(Any Two Channels at Equal Vertical Scale)
≥100:1 at ≤100 MHz and ≥30:1 at >100 MHz up to the rated bandwidth
Vertical System Digital Channels
  MSO4032 MSO4034 MSO4054 MSO4104
Input Channels 16 Digital (D15-D0)
Thresholds Per Channel Thresholds
Threshold Selections TTL, CMOS, User Defined
User-defined Threshold Range +5 to -2V
Maximum Input Voltage ±15V
Threshold Accuracy ±(100 mV + 3% of threshold setting)
Input Dynamic Range 6 Vpk-pk centered around the threshold
Minimum Voltage Swing 500 mV
Input Impedance 20 kW
Channel-to-Channel Skew 1 ns typical
Probe Loading 3 pF
Vertical Resolution 1 bit
Horizontal System Characteristics
  DPO4032
MSO4032
DPO4034
MSO4034
DPO4054
MSO4054
DPO4104
MSO4104
Maximum Sample Rate (all channels) 2.5 GS/s 2.5 GS/s 2.5 GS/s 5 GS/s
Maximum Record Length (all channels) 10 M points
Maximum Duration at Highest Sample Rate (all channels) 4 ms 4 ms 4 ms 2 ms
Timebase Range (s/div) 1 ns to 1,000 s 400 ps to 1,000 s
Timebase Delay Time Range –10 divisions to 50 s
Channel-to-channel Deskew Range ±100 ns
Timebase Accuracy ±5 ppm over any ≥1 ms interval
Horizontal System Digital Channels
  MSO4032 MSO4034 MSO4054 MSO4104
Maximum Sample Rate (Main) 500 MS/s
Maximum Record Length (Main) 10 M points
Maximum Sample Rate (MagniVu) 16.5 GS/s
Maximum Record Length (MagniVu) 10 K points centered on the trigger
Maximum Detectable Pulse Width 1.5 ns
Trigger System

Main Trigger Modes - Auto, Normal, and Single.

Trigger Coupling - DC, HF reject (attenuates >50 kHz), LF reject (attenuates <50 kHz), noise reject (reduces sensitivity).

Trigger Holdoff Range - 2 ns to 8 s.

Sensitivity

Internal DC Coupled - 0.4 div DC to 50 MHz increasing to 1 div at rated bandwidth.

External (auxiliary input) - 200 mV from DC to 50 MHz increasing to 500 mV at 250 MHz.

Trigger Level Range

Any Channel - ±8 divisions from center of screen.

External (auxiliary input) - ±8 V.

Trigger Modes

Edge - Positive or negative slope on any channel or front panel auxiliary input. Coupling includes DC, HF reject, LF reject, and noise reject.

Sequence (B-trigger) - Trigger Delay by Time - 4 ns to 8 s.  Or Trigger Delay by Events - 1 to 9,999,999 events.

Pulse Width - Trigger on width of positive or negative pulse that are >, <, = or ≠ a specified period of time.

Runt - Trigger on a pulse that crosses one threshold but fails to cross a second threshold before crossing the first again.

Logic - Trigger when any logical pattern of channels goes false or stays true for specified period of time. Any input can be used as a clock to look for the pattern on a clock edge. Pattern (AND, OR, NAND, NOR) specified for four input channels defined as High, Low or Don't Care.

Setup & Hold - Trigger on violations of both setup time and hold time between clock and data present on any of the input channels.

Rise/Fall Time - Trigger on pulse edge rates that are faster or slower than specified. Slope may be positive, negative or either.

Video - Trigger on all lines, odd, even, or all fields on NTSC, PAL, and SECAM video signals.

I2C (Optional) - Trigger on Start, Repeated Start, Stop, Missing ACK, Address (7 or 10 bit), Data, or Address & Data on I2C buses up to 3.4 Mb/s.

SPI (Optional) - Trigger on SS, MOSI, MISO, or MOSI & MISO on SPI buses up to 50 Mb/s.

CAN (Optional) - Trigger on Start of Frame, Frame Type (data, remote, error, overload), Identifier (standard or extended), Data, Identifier and Data, End of Frame or Missing ACK on CAN signals up to 1 Mb/s. Data can be further specified to trigger on ≤,<,=, >, ≥ or ≠ a specific data value. User-adjustable sample point is set to 50% by default.

RS-232 (optional) – Trigger on Tx start bit, Rx start bit, Tx end of packet, Rx end of packet, Tx data, Rx data.

Parallel (available on MSO models only) – Trigger on a parallel bus data value.

Acquisition Modes

Sample - Acquire sampled values.

Peak Detect - Captures narrow glitches at all real-time sampling rates.

Averaging - From 2 to 512 waveforms included in average.

Envelope - Min-max envelope reflecting Peak Detect data over multiple acquisitions.

Hi-Res - Real-time boxcar averaging reduces random noise and increases vertical resolution.

Roll - Scrolls waveforms right to left across screen at sweep speeds slower than or equal to 40 ms/div.

Waveform Measurements

Cursors - Waveform and Screen.

Automatic Measurements - 25, of which up to 4 can be displayed on screen at any one time. Measurements include Period, Frequency, Delay, Rise Time, Fall Time, Positive Duty Cycle, Negative Duty Cycle, Positive Pulse Width, Negative Pulse Width, Burst Width, Phase, Positive Overshoot, Negative Overshoot, Peak to Peak, Amplitude, High, Low, Max, Min, Mean, Cycle Mean, RMS, Cycle RMS, Area, and Cycle Area.

Measurement Statistics - Mean, Min, Max, Standard Deviation.

Reference Levels - User definable reference levels for automatic measurements can be specified in either percent or units.

Gating - Isolate the specific occurrence within an acquisition to take measurements, on using either the screen or waveform cursors.

Waveform Math

Arithmetic - Add, subtract, multiply, and divide waveforms.

Math Functions - Integrate, Differentiate, FFT.

FFT - Spectral magnitude. Set FFT Vertical Scale to Linear RMS or dBV RMS, and FFT Window to Rectangular, Hamming, Hanning, or Blackman-Harris.

Advanced Math - Define extensive algebraic expressions including waveforms, math functions, scalars, up to two user adjustable variables, and results of parametric measurements e.g. (Intg(Ch1-Mean(Ch1)) × 1.414 × VAR1).

Software

National Instruments SignalExpress™ Tektronix Edition – A fully interactive measurement software environment optimized for the 4000 Series, enables you to instantly acquire, generate, analyze, compare, import and save measurement data and signals using an intuitive drag-and-drop user interface that does not require any programming. Standard 4000 Series support for acquiring, controlling, viewing and exporting your live signal data is permanently available through the software. The Professional version adds additional signal processing, advanced analysis, mixed signal, sweeping, limit testing and user-defined step capabilities and is available for a 30-day trial period standard with each instrument.

OpenChoice® Desktop – Enables fast and easy communication between a Windows PC and the 4000 Series, via USB or LAN. Transfer and save settings, waveforms, measurements and screen images.

IVI Driver – Provides a standard instrument programming interface for common applications such as LabVIEW, LabWindows/CVI, Microsoft .NET and MATLAB.

Display Characteristics

Display Type - 10.4” (264 mm) liquid crystal TFT color display.

Display Resolution - 1,024 horizontal×768 vertical pixels (XGA).

Waveform Styles - Vectors, Dots, Variable Persistence, Infinite Persistence.

Input/Output Ports

CompactFlash Drive - Front panel access (Type 1).

USB 2.0 Full Speed Host Port - Supports USB mass storage devices and printers. Two ports available on rear-panel and one on front-panel.

USB 2.0 High Speed Device Port - Rear panel connector allows for control of oscilloscope via USBTMC or GPIB with a TEK-USB-488.

LAN Port - RJ-45 connector, supports 10/100Base-T.

XGA Video Port - DB-15 female connector, connect to show the oscilloscope display on an external monitor or projector.

Auxiliary Input - Front panel BNC connector. Input Impedance 1 MΩ. Max input 250 VRMS with peaks ≤±400 V.

Probe Compensator Output - Front panel pins. Amplitude 2.5 V. Frequency 1 kHz.

Trigger Out - Rear-panel BNC connector, provides a positive polarity pulse when the oscilloscope triggers.

Kensington Lock - Rear-panel security slot connects to standard Kensington lock.

Power Source

Power Source Voltage - 100 - 240 V ±10%

Power Source Frequency - 47 – 66 Hz (90 to 264 V), 360 – 440 Hz (100 to 132 V)

Power Consumption - 250 W maximum

Physical Characteristics

  Benchtop Configuration Rackmount Configuration
Dimensions mm in. mm in.
Height 229 9.0 221 8.7
Width 439 17.3 432 17.0
Depth 137 5.4 114 4.5
Weight kg lbs. kg lbs.
Net 5 11 9.1 20
Shipping 9.5 22 - -

General Characteristics

Cooling clearance - 2" (51 mm) required on left side and rear of instrument

Environmental

Temperature

Operating - 0 °C to +50 °C

Non-Operating - –20 °C to +60 °C

Humidity

Operating - High: 40 °C to 50 °C, 10% to 60% Relative Humidity.

Low: 0 °C to 40 °C, 10% to 90% Relative Humidity.

Non-Operating - High: 40 °C to 60 °C, 5% to 60% Relative Humidity.

Low: 0 °C to 40 °C, 5% to 90% Relative Humidity.

Altitude

Operating - 9,843 feet (3,000 m).

Non-Operating - 39,370 feet (12,000 m).

Random Vibration

Operating - 0.31 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total

Non-Operating - 2.46 GRMS from 5 to 500 Hz, 10 minutes each axis, 3 axes, 30 minutes total

Regulatory

Electromagnetic Compatibility - 89/336/EEC Safety - UL61010-1, Second Edition; CSA61010-1 Second Edition, EN61010-1: 2001; IEC 61010-1: 2001.

 

DPO 4000 Series Oscilloscopes     MSO 4000 Series Oscilloscopes

 

Tektronix Catalog   |   Tektronix Price List

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved