<< Tektronix Accessories | << Tektronix Oscilloscope Probes | << Tektronix Current Probes
<< Tektronix Probes & Accessories Price List
Tektronix TCP404XL AC/DC Current Probe, DC to 2 MHz;
500 A DC (750 A DC Derated With Duty Cycle)
(Includes the required TCPA400 Amplifier)
|
 
TCPA300 & TCPA400 AC/DC Current Measurement Systems Shown
|
Features & Benefits
- AC/DC Measurement Capabilities
- DC – 2 MHz, 750*1
A DC (500 A DC Continuous)
- Automatic Scaling and Units*2
– Oscilloscope On-Screen Readout of Magnitude and Amps Reduces
Measurement Errors with No More Hand Calculations
- AC/DC Input Coupling
- Low Insertion Impedance Reduces Device Under Test Loading
- Split-Core Construction Allows Easy Circuit Connection
- Status Indicators Provide Visual Operating Status and
Notification of Potential Error Conditions – Degauss, Probe Open,
Overload, Not Terminated into 50 Ohm, Noncompatible Probe Type
- Lower DC Drift and Noise Allows Improved Low-level Current
Measurements
- Certified for use in U.S., Canada, and Europe. Complies
with applicable IEC standards.
Applications
- Development and Analysis Solutions for Designers, Installers, and
Service Personnel in Telecomm, Data Comm, Computer, and Semiconductor
Power Electronics Environments For:
- Power Supplies (Switching and Linear)
- Semiconductor Devices (SCRs, IGBTs, MOSFETs, CMOS, BJTs)
- Power Inverters/Converters
- Electronic Ballasts
- Industrial/Consumer Electronics
- Mobile Communications (Phone, Satellite, Relay Stations)
- Motor Drives
- Transportation Systems (Electronic Vehicles, Electric Trains,
Locomotives, Avionics)
Increased performance and simplicity
Tektronix TCP300 and TCP400 series AC/DC Current Measurement family is
a highly advanced current measurement system for today's current
measurement needs. The split-core probes incorporate both transformer and
Hall-effect technology to provide broad bandwidth measurement capabilities
from DC up to 100 MHz at current levels as low as 1 milliamp to thousands
of amps (when used with CT4 current transformer). When connected to
Tektronix Oscilloscopes with TEKPROBE Level II, TekConnect (w/ TCA-BNC) or
TekVPI (w/ TPA-BNC) interfaces, current measurements and calculations are
simple and easy.
Additional measurement power is available with add-on software such
as the TDSPWR2 power measurements package. With all this measurement
power, current measurements have never been easier.
*1 Derated with Duty Cycle *2
Requires a TDS TEKPROBE Oscilloscope or a TekConnect Oscilloscope with
TCA-BNC
Meets today's AC/DC current measurement applications
The TCPA300 amplifier, when used with
TCP312,
TCP305, or
TCP303
probes, provide a wide range of current measurement capability and spans
the gap between low-level milliamp measurements to very high current
levels. These three probes provide current measurement capabilities of
30 A, 50 A, and 150 A DC continuous. For even higher current levels, the
TCPA400 amplifier with the TCP404XL current probe measures 500 A DC
continuous and 750 A DC continuous, derated
with duty cycle.
Higher frequency performance is available with the
TCP312 w/TCPA300 providing ≥100-MHz bandwidth and a maximum current of 30 A
DC.
Measurement errors and manual calculations are now a thing of
the past
With this new series of current measurement tools, automatic
control and on-screen scaling and units is provided for users of
Tektronix TDS3000, TDS500, TDS600, TDS700, TDS5000, TDS6000, and
TDS7000B series oscilloscope systems (the DPO3000, DPO4000, and
DPO7000 series oscilloscopes, the TPA-BNC adapter is required).
The TCP300/TCP400 Current Measurement Systems seamlessly
integrate with your TDS series oscilloscope and the TDSPWR2 software
package to easily make those time-consuming power measurements and
calculations for you.
Even non-TEKPROBE systems can use the TCPA 300/400 series to make
proper current measurements by simply multiplying the measured
output voltage on the oscilloscope by the TCPA 300/400 series range
setting.
Specifications
|
|
TCP312 w/ TCPA300 |
TCP305 w/ TCPA300 |
TCP303 w/ TCPA300 |
TCP404XL w/ TCPA400 |
|
Bandwidth (-3 dB) |
DC – 100 MHz |
DC – 50 MHz |
DC – 15 MHz |
DC – 2 MHz*1
|
|
Risetime |
≤3.5 ns |
≤7 ns |
≤23 ns |
≤175 ns |
|
Maximum Current Ratings: |
|
High-Current Sensitivity Range |
10 A/V Range |
10 A/V Range |
50 A/V Range |
1 A/mV Range |
|
DC (continuous) |
30 A |
50 A |
150 A |
500 A (750 A*2) |
|
RMS (sinusoidal) |
21.2 A |
35.4 A |
150 A |
500 A |
|
Peak Pulse |
50 A |
50 A |
500 A |
750 A |
|
Low-Current Sensitivity Range |
1 A/V Range |
5 A/V Range |
5 A/V Range |
N/A |
|
DC (continuous) |
5 A |
25 A |
25 A |
|
|
RMS (sinusoidal) |
3.5 A |
17.7 A |
17.7 A |
|
|
Peak Pulse |
50 A |
50 A |
500 A |
|
|
DC Accuracy (Operating temp 0 °C to 50 °C) |
±3% of reading |
±3% of reading |
±3% of reading (10 °C to 50 °C); +3%/-6% of reading (0 °C to <10 °C) |
±3% of reading |
|
DC Accuracy, Typical (Operating temp 23 °C ±5 °C) |
±1% of reading |
±1% of reading |
±1% of reading |
±1% of reading |
|
Nominal
|
|
Maximum Bare Wire Voltage |
For Use With Insulated Wires Only |
600 VRMS CAT I & II; 300 VRMS CAT III |
|
Lowest Measurable Current (at ±3% accuracy at DC) |
1 mA |
5 mA |
5 mA |
1 A |
|
Scope Set To 1 mV/div and 20 MHz BW Limited
|
Scope Set To 1 mV/div and 20 MHz BW Limited
|
|
Insertion Impedance (See Curves Below) |
0.08 Ω at 1 MHz ;0.15 Ω at 10 MHz; 0.27 Ω at 50 MHz; 0.7 Ω at 100 MHz |
0.035 Ω at 1 MHz; 0.12 Ω at 10 MHz; 0.4 Ω at 50 MHz |
0.01 Ω at 1 MHz; 0.025 Ω at 5 MHz; 0.1 Ω at 15 MHz |
0.1 mΩ at 10 kHz; 0.6 mΩ at 100 kHz; 8 mΩ at 1 MHz; 16 mΩ at 2 MHz |
| Typical |
|
Maximum Amp-Second Product (Based on Amplifier Range setting) |
50 A*μS – 1 A/V; 500 A*μS – 10 A/V |
500 A*μS – 5 A/V; NA – 10 A/V |
3,000 A*μS – 5 A/V; 15,000 A*μS – 50 A/V |
NA – 1 A/mV |
|
AC-Coupling Low-Frequency Bandwidth (Low Pass - 3 dB point) |
<7 Hz |
|
Displayed RMS Noise (at 20-MHz Bandwidth Limit) |
≤250 μARMS
|
≤1.25 mARMS
|
≤2.5 mARMS
|
≤250 mARMS
|
|
Signal Delay (Delay to Output BNC) |
17 ns |
19 ns |
40 ns |
80 ns |
|
Inputs (probe amplifier) |
1 |
|
Probe Open Indicator |
YES |
|
Overload Indicator |
YES |
|
Termination Indicator |
YES |
|
Noncompatible Probe Indicator |
YES |
|
Safety Certifications
|
|
U.S. NRTL Listing |
UL3111-1 (Amplifier) |
UL3111-2-032 ; UL3111-1 (Probe and Amplifier) |
|
Canadian Certification |
CAN/CSA C22.2 No.1010.1; (Amplifier) |
CAN/CSA C22.2 No.1010.2.032; CAN/CSA C22.2 No.1010.1;(Probe and Amplifier) |
|
European Union Compliance |
EN61010-1/A2; (Amplifier) |
EN61010-1/A2; EN61010-2-032; EN61010-1/A2; (Probe and Amplifier) |
|
Other |
IEC61010-1/A2 (Amplifier) |
IEC61010-2-032; IEC61010-1/A2; (Probe and Amplifier) |
|
Electromagnetic Compatibility |
EC Council Directive 89/336/EEC, FCC Part 15, Subpart B Class A, AS/NZS
2064.1/2. |
|
Power Requirements (TCPA300/TCPA400 Amplifiers) |
90 V to 264 V; 47 to 440 Hz; 50 W; Maximum CAT II (Auto Switch) |
|
Power Requirements (Probes) |
Requires TCPA300 Amplifier |
Requires TCPA400 Amplifier |
|
Probe Model |
TCP312 |
TCP305 |
TCP303 |
TCP404XL |
|
Warranty |
1 Year |
|
Probe Mechanical Characteristics
|
|
Probe Cable Length |
1.5 meters (60 inches) |
2 meters (78.7 inches) |
8 meters (315 inches) |
|
Probe Jaw Size (Max Conductor Size) |
3.8 mm (0.15 inches) |
21 mm × 25 mm (0.83 × 1.0 inches) |
|
Length |
20 cm (7.87 inches) |
26.8 cm (10.55 inches) |
26.8 cm (10.55 inches) |
|
Width |
1.6 cm (0.625 inches) |
4.1 cm (1.60 inches) |
4.1 cm (1.60 inches) |
|
Height |
3.2 cm (1.25 inches) |
15.6 cm (6.13 inches) |
15.6 cm (6.13 inches) |
|
Weight |
0.15 kg (0.33 lbs.) |
0.66 kg (1.45 lbs.) |
0.88 kg (1.90 lbs.) |
|
TCPA300 and TCPA400 Mechanical Characteristics
|
|
Length |
17.3 cm (6.8 inches) |
|
Width |
16.7 cm (6.6 inches) |
|
Height |
9.14 cm (3.6 inches) |
|
Weight |
1.14 kg (2.5 lbs.) |
|
Environmental Characteristics
|
|
Temperature - Operating |
0 °C to +50 °C (32 °F to 122 °F) |
|
Temperature - Nonoperating |
−40 °C to +75 °C (−40 °F to 167 °F) |
|
Humidity – Operating |
5% to 95% R.H. to +30 °C (86 °F); 5% to 85% R.H. +30 °C to +50 °C (86 °F to 122 °F) |
|
Humidity – Nonoperating |
5% to 95% R.H. to +30 °C (86 °F); 5% to 85% R.H. +30 °C to +75 °C (86 °F to 167 °F) |
|
Altitude - Operating |
2000 m (6800 ft.) maximum |
|
Altitude - Nonoperating |
12,192 m (40,000 ft.) maximum |
|
|
|
| Ordering Information |
| Model |
Description |
Buy Online |
Price |
| TCP404XLTCPA400 |
Tektronix TCP404XL Current Probe with TCPA400 Amplifier, 500 A, 2 MHz AC/DC |
 |
Sale $5,211.00 (Reg. $5,790.00) |
| Included Accessories |
| |
- Current Probe Includes:
- (1) AC/DC Current Probe
- (1) Instruction Sheet
- (1) Probe Cover
- (1) Certificate of Traceable Calibration
- Current Probe Amplifier Includes:
- (1) AC/DC Current Probe Amplifier
- (1) Instruction/Service Manual
- (1) TEKPROBE Interface Cable
- (1) Male to Male BNC cable (50 Ω)
- (1) 50 Ω feed-through termination
- (1) Certificate of Traceable Calibration
|
Included |
| Recommended Accessories |
| 016192300 |
Tektronix 016-1923-00 Cover, Small Probe Protective; (for TCP305,
TCP312, A6302, A6302XL, A6312, TCP202) |
Please Call |
| 016192400 |
Tektronix 016-1924-00 Cover, Large Probe Protective; (for TCP303,
TCP404XL, A6303, A6303XL, A6304XL) |
Please Call |
| 016192200 |
Tektronix 016-1922-00 Case, Transit; Current Measurement Systems |
Please Call |
| 011004902 |
Tektronix 011-0049-02 50 Ω Feed Through Termination |
Please Call |
| 012011700 |
Tektronix 012-0117-00 50 Ω BNC To BNC Coaxial Cable |
Please Call |
| 012160500 |
Tektronix 012-1605-00 TEKPROBE Interface Cable, TCPA300 or
TCPA400 Amplifier to TDS series oscilloscopes |
Please Call |
| 067147800 |
Tektronix 067-1478-00 - Power Measurements Deskew Fixture, For
TCP202, TCP305, TCP312, TCP303, A6302, A6312, A6303, Series
Probes |
Please Call |
| TDSPWR2 |
Tektronix TDSPWR2 - Power Measurement and Analysis Software for
TDS5000 and TDS7000 Series Oscilloscopes; To order when purchasing oscilloscope
use models: TDS5052PW2, TDS5054PW2, TDS7054PW2,TDS7104PW2.
To order as an upgrade use: TDS5UP PW2, TDS7UP PW2
|
Please Call |
|