Test Equipment Depot - 800.517.8431 McAfee Secure sites help keep you safe from identity theft, credit card fraud, spyware, spam, viruses and online scams

Picotest U6200A
400 MHz Universal Frequency Counter

  • Measuring frequency over 400MHz within standard 6GHz Channel 3
  • 12 digits resolution with 1s gate time
  • 40ps time domain function resolution
  • Great Features for Universal Purposes
  • Full Math Functions & Easy Operation Panel
  • Fast Measurement & Special Application
  • Free Software & Familiar SCPI Commands

Click here for a larger image - Picotest U6200A Universal CounterClick here for a larger image - Picotest U6200A Universal Counter
Picotest U6200A Shown

Product Information
Datasheet 379 KB
Manual 1.25 MB
Specification
Charts
1.33 MB
Comparison Guide

HTML

PDF Help
 
The Picotest U6200A universal counter, whose production procedures conform to ISO 9004, has frequency resolution of 12 digits per second, 40 ps time interval resolution and a complete set of test and analysis features. The standard U6200A’s CH3 comes with the range from 375 MHz to 6GHz and the standard CH 1 & 2 from 1 mHz to 400 MHz.

The Picotest U6200A also provides great features including Frequency & Ratio (11Digits/Sec.), Time interval, Period (2.5 ns to 1000s), Duty Cycle, Pulse Width, Rise/Fall Time, Peak Volts (100 Hz~300 MHz), Phase, Totalize, Temperature Stability (< 1 PPM), Aging Rate (< 2 PPM per year), time base reference channel and complete Front-End Isolation. Moreover, it offers 20 memories for storing frequently-used operations.

The Picotest U6200A offers built-in statistics and math functions. Users can do measurements, simultaneously measure, count in mean, min/max, delta & standard deviation. And scale & offset can be easily used for compensation purpose according to users’ applications. In addition, to reach these measurements, user can easily use the numeric buttons to define settings. Moreover, the U6200 also provides users the visible light buttons as functions are working.

The U6200A supports real-time digital signal processing technology, which is applied to analyze data as simultaneously getting new readings and speeding measurement. The “Limit Modes” is worth to be mentioned, i.e. users can set margin according to their specific measurements, and via Go-On or Stop and USB Output settings, the U6200A can be keep or stop measuring as a limit is exceeded, and generate an output signal to trigger external devices.

Users can get data logs by PC software (Microsoft Excel®) via a USB, or an optional GPIB interface. Furthermore, U6200A supports a web server function, so users can easily control it via a LAN interface (Figure 10) by inputting an Ethernet address on web browsers. In addition, by the SCPI commands compatible to Agilent 53132A, the Picotest U6200A can provide familiar syntax string for users’ applications. For more command information, please refer to the U6200A User’s Manual in Chapter 7.
 
Specifications
Channel 1 & 2 Input Specifications
DC Coupled 1mMHz
AC Coupled 200 KHz to 400 MHz (50Ω)
30 Hz to 400 MHz (1Ω)
FM Tolerance FM Tolerance: 25%
Voltage Range and Sensitivity
1mH to 225 MHz 20 mVrms to ± 5 V ac + dc
(75 mVrms with optional rear connectors)
225mH to 400 MHz 30 mVrms to ± 5 V ac + dc
(75 mVrms with optional rear connectors)
Channel 1 & 2 Input Characteristics
Impedance (ATTx1, 1 MΩ Capacitance) 24pF
(ATTx10, 1 MΩ Capacitance) 15pF
Coupling AC or DC
Low-Pass Filter 100 KHz (or disabled)
-20 dB at > 1 MHz
Input Sensitivity Selectable between Low, Medium (default), or High Medium is approximately 1.35x High Sensitivity, low is approximately 1.7x High Sensitivity
Voltage Range and Sensitivity (Single-Shot Pulse)
1.5ns to 10ns Pulse Width 80 mVpp to 10 Vpp
(150 mVrms with optional rear connectors)
225mH to 400 MHz 50 mVrms to 10 Vpp
(150 mVrms with optional rear connectors)
Trigger Level (ATT x 1)
Range ±5.125V
Accuracy ±(15mV + 1% of Trigger Level)
Resolution 2.5mV
ATT x 10 Range x10
Trigger Slope Positive or Negative
Auto Trigger Level Range 0 to 100% in 1% Steps
Frequency Peak Voltage Fast Mode > 10KHz
Peak Voltage Slow Mode > 100Hz
Amplitude > 100mVpp (No Amplitude Modulation)
Damage Level
DC ~400MHz 50Ω 12 Vrms
0 to 3.5 lHz to 1MΩ 350 V dc + ac pk
0 to 3.5 lHz to 100KHz, 1MΩ 350 V dc + ac pk linearly derated to 12 Vrms
100KHz to 400MHz, 1MΩ 12 Vrms
Attenuator
Voltage Range x10
Trigger Range x10
Specification Chart

Click here for additional information (PDF) >>

 
General Specifications
Function Description
Power Supply Voltage 100V/240V ± 10% 50Hz ~ 60Hz ± 10%
100V/240V ± 10% 400Hz ± 10%
Power Requirements 50VA Maximum
Operating humidity Maximum relative humidity 80% for temperature up to 30°C decreasing linearly to 50% relative humidity to 40°C
Operating Environmental 0 to 55°C
Storage Temperature -40°C to 70°C
Operating Altitude Up to 2000m
Bench Dimensions
(WxHxD)
8.27in. x 3.35in. x 13.8 (210mm x 85mm x 350mm)
Weight 7.1 LBs
Safety IEC61010-1: 2001/EN61010:1:2001 (2nd Edition)
EMC EN61323, IE 61000-3, IE61000-4
Warm-up Time 1 Hour
Warranty

1 Year

 

Ordering Information

Model

Description

Buy Online

Price
U6200A  Picotest U6200A Universal Counter Buy Now $1,750.00 
U6200AGC  Picotest U6200AGC Universal Counter with built-in GPIB Card Buy Now $1,845.00 
Included Accessories
 
  • 1 - Power Cord

Included

Picotest Catalog | Picotest Price List

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved