General Specifications |
|
Measurement Range
|
-
Luminance 0.1 to 300 cd/m2 (0.03 to
100 ftL)
-
x and y values: 0.000 to 0.800
-
Correlated Color Temperature: 1.900 K
CCT to 12.000 K CCT
|
|
Accuracy
|
-
The following specifications apply to
a measurement with an illuminant D6500 standard monitor1* at a
luminance level of 10 cd/m2 (3ftL) and at an ambient temperature
between +15°C (+59°F) and +30°C (86°F)
-
xy coordinates: Better than ± 0.002
-Repeatability: Better than ± 0.002
-Luminance (Y): Better than ±2% ± 1 digit
-Repeatability: Better than ± .03%, ± 1 digit
-
XYZ/RGB bars: Better than ±1%
-Repeatability: Better than ±1%
Luminance (Y): Better than ±2%, ± 1 digit
-
Measuring rate: 3 to 10
measurements/second
-
Settling time: Within 1% after 1
second
-
Correlated color temperature: ±50 K CCT
1*Calibration of the standard monitor is
traceable to international light standards
|
|
Memory
|
-
References: The white references are
stored in the controlling computer. The number of white references are
limited only by the computer. Legal values for x and y are between
0.200 and 0.600
-
CRT/Phosphors: 30 different phosphors
may be stored in the color sensor. Additional phosphors may be stored
in the PC. Phosphors may be stored vial the "Learn phosphor"
function. (only applicable when using the RGB bar display)
-
Set-ups: An unlimited number of
measuring setups may be stored in the C. A measuring setup includes
the display mode, CRT/phosphor, white reference, and measuring units
-
Measurements: An unlimited number of
measurements may be stored for later investigation and statistical
analysis
Stored measurements may also be used as white references
|
|
Factory Programming
|
-
White ref.:
-D6500 (x=o.313, y=0.329)
-3200 K (x=0.423, y=0.399)
-9300 K (x=0.285, y=0.293)
-
Other white references may be stored
using the "Learn white reference" function, as measurements
or put directly in the file containing the white reference
|
|
Power
|
-
The color sensor is powered from a
power supply box inserted between the PC and the color sensor
-
Mains consumption:
85-250 V AC 15 VA. 48 to 65 Hz
|
|
Environmental Conditions
|
-
Operating temperature:
10°C to +40°C (+50°F to +104°F) (non-condensing)
-
Storage temperature:
-10°C to +70°C (-14°F to 158°F)
-
Vibration: IEC 68-2-6 F; 5-50-200 Hz
0.7 mm p-p 50 m/s2; 1 oct/min; 3x30 min
-
Bump: IEC 68-2-29 part 2; 350 m/s2
1000 bumps in 3 directions
-
Repetitive bump: 120 m/s2
|
|
Mechanical Data |
|
Dimensions
|
-
Color Sensor:
Diameter of house: 108 mm (4.25")
Diameter of rubber shadow pad: 120 mm (4.75")
Height: 70 mm (2.75")
Weight: 250 g (0.55 lbs)
-
Power supply box:
Size: 185x95x65 mm (7.3"x3.8"x2.6")
|
|
Hardware Requirements
|
-
IBM PC AT/386/486/Pentium, PS/2 or
compatible
-
512 bytes RAM
-
One floppy disk drive (5-1/4" or
3-1/2")
-
MS-Dos, PC-DOS or DR-DOS (version 3.3
or higher)
-
For full use of the display facilities
a 100% compatible IBM EGA/VGA graphic adapter is recommended
-
The software can be run from a floppy
disk but a hard disk is recommended
|