| Specifications |
| Electrical supply range |
120 V a.c. ±10% |
| Supply measurement |
Voltage |
±2% ±1 V |
| Frequency |
±1% ± 0.1 Hz |
| Bond test |
Open circuit voltage |
9 V a.c. ± 10% ± 0.1 V |
| 10 A bond test current |
10 A rms ±25% ± 0.5 A into 0.1 Ω |
| 25 A bond test current |
25 A rms ± 20% ± 0.5 A into 0.1 Ω |
| Earth bond resistance accuracy |
±5% ±3 digits (0 to 0.49 Ω) ±5% ±5 digits
0.5 to 1.99 Ω) |
| Earth bond resistance resolution |
10 mΩ (0 to 1.99 Ω) |
| Display range |
0 to 1.99 Ω |
| Bond test nulling |
Up to 1.00 Ω |
| Adjustable test duration |
User selectable from 1 sec to 20 sec |
| Insulation test |
Insulation test voltage |
250 V d.c. –0%/+25% open circuit 500 V d.c.
–0%/+25% open circuit (500 V d.c. across 0.5 MΩ) |
| Insulation resistance accuracy |
±2% ±5 digits (0 to 19.99 MΩ) ±5% ±10
digits (20 to 99.99 MΩ) |
| Insulation resistance resolution |
0.01 MΩ (0.10 to 99.99 MΩ) |
| Display range |
0 to 99.99 MΩ |
| Test duration |
User selectable from 1 sec to 1 minute |
| Continuity test |
Continuity test voltage |
4.0 V d.c. -0% +10% (open circuit) |
| Continuity test current |
200 mA -0% +10% ±5 mA (into 2 Ω load) |
| Continuity resistance accuracy |
±5% ±3 digits (0 to 0.49 Ω) ±5% ±5 digits
(0.5 to 19.99 Ω) |
| Continuity resistance resolution |
10 mΩ (1 to 19.99 Ω) |
| Display Range |
0 to 19.99 Ω |
| Continuity test nulling |
up to 9.99 Ω |
| Test duration |
User selectable from 1 sec to 20 sec |
| Substitute leakage test |
Test voltage |
40 V a.c. ±10% |
| Test frequency |
Nominal mains 50/60 Hz |
| Leakage current accuracy |
±5% ±5 digits |
| Leakage current resolution |
0.01 mA |
| Display range |
0 to 19.99 mA |
| Test duration |
User selectable from 1 sec to 1 minute |
| Differential leakage current |
Test voltage |
Nominal mains 120 V a.c. |
| Test frequency |
Nominal mains 60 Hz |
| Accuracy |
±5% ±5 digits |
| Resolution |
0.01 mA |
| Display range |
0 to 19.99 mA |
| Touch current Test |
Test voltage |
Nominal mains 120 V a.c. |
| Test frequency |
Nominal mains 60 Hz |
| Touch current Accuracy |
±5% ±5 digits |
| Touch current resolution |
0.01 mA |
| Display range |
0 to 10 mA |
| Test duration |
User selectable from 1 sec to 5 sec |