<< Instek LCR Meters |
Instek LCR-8101G 20Hz ~ 1MHz Precision LCR Meter
|
| |

Instek LCR-8101G Shown
|
|
|
- Wide Test Frequency 20Hz ~ 10/5/1MHz
- 0.1% Basic Accuracy & 6 Digits Measurement Resolution
- Large LCD Display with Intuitive Interface
- Full Measuring Functions with DUT V/I Monitor
- PASS/FAIL Function ( abs, % or delta) with Judgment Alarm
- DC Resistance Measurement
- Multiple Step Mode
- Standard RS-232C/GPIB Interface
|
| |
The Instek LCR-8101G LCR meter, with test
frequency up to 1MHz, provides accuracy, versatility and high
resolution for a wide range of component measurements, even including DC
resistance measurement and Voltage/Current monitoring. The Multi-Step
function allows on-screen programming of customized measurement sequence
with Pass/Fail indication. Each program includes 30 test steps and each
test step can be set with selected parameters and test limits. Under
Multi-Step operation, a tedious work routine can be done step by step
automatically just at a press of a button. With Graph Mode, Instek LCR-8101G display the component impedance
response either over a wide range of test frequency sweep or over a wide
range of test voltage sweep in a graph chart. This gives an analysis
result of either impedance vs. frequency or impedance vs. applied
voltage all at a glance. GPIB and RS-232C interfaces are available as
standard for instrument control and test result display on the PC. The
rich features of LCR8101G easily make your measurement tasks
done at a very competitive price.
The Instek LCR8101G is
designed to perform precision impedance measurements over a wide
frequency range of 20Hz~1MHz for LCR-8101G.The instrument is capable of measuring 11
different parameters with0.1% basic accuracy, which meets the precision
measurement requirements of components and modules used in the RF
circuits. The large LCD display with single-layer operation menu of
LCR-8101G Series provides users with ultimate convenience to plug and
play without much learning time.
|
In the Pass/Fail test, primary
parameter measurement result is compared with user-defined limits and
the pass or fail result is then displayed. Three methods, including
absolute limit, percentage and delta limit, are available for Hi(high)
and Lo(low) limit setting based on the nominal test value. The Pass/Fail
test checks whether the parameter measurement result sits within the
limits, then display "PASS" for within the limits, or "LO" for lower
than the low limit, or "HI" for higher than the high limit. A scale and
bar for displaying measurement result is shown at the center of the
screen to give a graphical identification, which greatly reduces
operator's load in a long time inspection work. This scale and bar also
facilitate the adjustment of the variable components. Either Pass or
Fail result can be set with a buzzer alarm, which makes component or
material sorting easy with sound identification.
|
The Multi-Step
mode is capable of running a series of measurements of a component at a
number of user-defined steps in sequence automatically. Total 64
programs can be saved into the non-volatile memory, and each program
contains up to 30 test steps. The parameter and Hi/Lo limits can be set
respectively for each test step. After a program being properly edited,
the instrument can run through all the measurement steps either at a
press of the button under the Manual Trigger selection, or automatically
run the program by detecting the connection of a DUT under the Auto
Trigger selection. When all the test steps are completed, the screen
shows the measurement reading of the parameter being selected for each
step with Pass, HI, or LO measurement result.
|
| The graph function
shows the component characteristics in visual manner. Either voltage
sweep or frequency sweep can be selected for horizontal scale. Just
select the parameter, and set the start/stop voltage or the start/stop
frequency of the sweep, LCR-8101G will run through the sequence of
measurements and show the results on a graph. This graphical parameter
measurement performs the characteristic verification of components and
materials over the response to the changes in AC test frequency or AC
test voltage without the need of complex programming of an external
controller. When the graph gets out of the vertical range, the LCR8110G Series can automatically adjust the scale to get full test information.
On the graph the marker operation is available for detailed observation. |
| |
| Specifications |
| Test Frequency |
20Hz ~ 1MHz, 5 Digits, ±0.005% |
| Input Impedance |
100Ω |
| Basic Accuracy |
±0.1% (R, Z, X, G, Y, B, L, C) |
| Test Speed |
AC |
Max: 75mS Fast: 150mS Medium: 450mS Slow: 600mS |
| DC |
Max: 30mS Fast: 60mS Medium: 120mS
Slow: 900mS |
| Test Signal Levels |
≤ 3MHz: 10mV~2Vrms, 1mV/Step or 10mV/Step, 2%±5mV
> 3MHz:10mV~1Vrms, 1mV/Step or 10mV/Step, 2%±5mV |
| Measured Range |
R, Z, X: 0.1mΩ ~ 100MΩ
G, Y, B: 10nS ~ 1000S
L: 0.1nH ~ 100kH
C: 0.01pF ~ 1F
D: 0.00001 ~ 9.9999
Q: 0.1 ~ 9999.9
θ: -180° ~ +180°
|
| Measurement Parameters |
Impedance (Z), Phase Angle (θ), Inductance (L), Capacitance (C),
AC Resistance (Rac), Quality Factor (Q), Dissipation Factor (D),
Admittance (Y), Conductance (G), Reactance (X), Susceptance (B),
DC Resistance (Rdc) |
| Series or Parallel equivalent circuit |
C + R, C + D, C + Q, L + R, L + Q, L + D |
| Series Equivalent Circuit Only |
X + R, X + D, X + Q |
| Parallel Equivalent Circuit Only |
C + G, B + G, B + D, B + Q, B + R, L + G |
| Polar Form |
Z + Phase Angle, Y + Phase Angle |
| LCD Display |
320 x 240 Dot-Matrix |
| Power Source |
AC 115V(+10% / -25%), 230V (+15% / -14%) (Selectable), 50/60Hz |
|
| |
| Ordering Information |
| Model |
Description |
Buy Online |
Price |
| LCR8101G |
Instek LCR-8101G Precision LCR Meter, 20 Hz - 1 MHz |
 |
$6,765.00 |
| Included Accessories |
| |
- 1 - Operation Manual
- 4 - Power Cord
- 1 - LCR-12 Kelvin Clip Test Lead
|
Included |
| Optional Accessories |
| Model |
Description |
Buy Online |
Price |
| LCR09 |
Instek LCR-09 SMD/Chip Test Fixture for LCR-816/817/819/821 |
 |
$262.00 |
| GRA404 |
Instek GRA-404 Rack Mount for GSP-827 |
 |
$250.00 |
| GTL232 |
Instek GTL-232 RS-232 Cable for GDS, GRS, SFG, LCR, PSM |
 |
$10.00 |
|