| Specifications |
| AC Withstanding |
| Output-Voltage Range
|
0.100kV~ 5.000kV ac
|
| Output-Voltage Resolution |
2V/step |
| Output-Voltage Accuracy
|
± (1% of setting + 5V) [no load]
|
| Maximum Rated Load |
200 VA (5kV/40mA) |
| Maximum Rated Current |
40mA (0.5kV< V ≤ 5kV)
10mA (0.1kV≤ V ≤0.5kV)
|
| Output-Voltage Waveform |
Sine wave
|
| Output-Voltage Frequency |
50 Hz / 60 Hz selectable
|
| Voltage Regulation |
± (1% of reading + 5V) [full load → no load] |
| Voltmeter Accuracy
|
± (1% of reading + 5V) |
| Current Measurement Range |
0.001mA~40.0mA |
| Current Best Resolution |
0.001mA / 0.01mA /0.1mA |
| GND |
RETURN/GUARD |
| DC Withstanding |
| Output-Voltage Range
|
0.100kV~ 6.000kV dc
|
| Output-Voltage Resolution |
2V/step |
| Output-Voltage Accuracy
|
± (1% of setting + 5V) [no load]
|
| Maximum Rated Load |
50W VA (5kV/40mA) |
| Maximum Rated Current |
10mA (0.5kV< V ≤ 6kV)
2mA (0.1kV≤ V ≤0.5kV)
|
| Voltage Regulation |
± (1% of reading + 5V) [full load → no load] |
| Voltmeter Accuracy |
± (1% of reading + 5V) |
| Current Measurement Range |
0.001mA~10.0mA |
| Current Best Resolution |
0.001mA / 0.01mA /0.1mA |
| GND |
RETURN/GUARD |
| Insulation Resistance |
| Output-Voltage
|
50V~1000V dc
|
| Output-Voltage Resolution |
50V/step |
| Output-Voltage Accuracy
|
± (1% of setting + 5V) |
| Resistance Measurement Range
|
1MW~ 9500MW |
| Test Voltage |
50V≤ V <500V 1~50MΩ:±(5% of reading + 1 MΩ)
51~2000MΩ:±(10% of reading + 1 MΩ)
|
500V≤ V ≤1000V 1~500MΩ:±(5% of reading + 1 MΩ)
501~9500MΩ:±(10% of reading + 1 MΩ)
|
| GND |
RETURN/GUARD |
| Ground Bond |
| Output-Current
|
03.00A~30.00A ac |
| Output-Current Resolution |
0.01A |
| Output-Current Accuracy
|
3A≤I≤8A:±(1% of reading + 0.2A)
8A<I≤30A:±(1% of reading + 0.05A) |
| Test-Voltage |
6Vac max (open circuit) |
| Test-Voltage Frequency |
50Hz/60Hz selectable |
| Resistance Measurement Range |
0.1mW~ 650.0mW |
| Resistance Measurement Resolution
|
0.1mW |
| Resistance Measurement Accuracy |
±(1% of reading + 2 mW) |
| GND |
RETURN/GUARD |
| Memory |
| Single Step Memory |
MANU:100 blocks |
| Automatic Testing Memory |
AUTO:100 blocks, manu per auto:16
|
| General |
| Display |
240 x 64 Ice Blue Dot matrix LCD |
| Power Source |
AC 100V/120V/220V/230V ± 10%, 50/60Hz |