Test Equipment Depot - 800.517.8431 McAfee Secure sites help keep you safe from identity theft, credit card fraud, spyware, spam, viruses and online scams
<< Hioki Oscilloscopes
<< Hioki Input Modules Index

Hioki 8958 16 Channel Scanner Input Module
for the Hioki Memory HiCorder

  • Used with the Hioki  8860-50 and 8861-50 Memory Hicorder

Click here for larger image - Hioki 8958 16 Channel Scanner Unit for the Hioki 886050 or 886150Click here for larger image - Hioki 8958 16 Channel Scanner Unit for the Hioki 886050 or 886150
Hioki 8958 Shown

Product Information

Datasheet 2.6 MB

PDF Help

 

Specifications

Specifications

Description
Measurement Functions Number of Channels 16
Type of Measurement Voltage Measurements, Temperature measurement with with Thermocouple
Input Connectors Voltage input/
Thermocouple
input
screw-type terminal strip, recommended wire diameter Recommended cable for Screw-type Terminal Strip: singlewire: 0.14 to 1.5 mm2, braided wire 0.14 to 1.0 mm2 (conductor wire diameter min. 0.18 mm), AWG 26 to 16
detachable terminal block (with cover)
Input impedance 1 MΩ, 850 kΩ with line fault detection ON, Max. rated voltage to earth: 33 Vrms or 70 V DC
Voltage Measurement Range 5m, 50m, 500m, 2 V/division, 4 ranges, full scale: 20 divisions
measurement range: ±100% of full scale, digital filter: 50/60/10 Hz, measurement resolution 1/1600 of measurement range (using 16-bit A/D conversion; installed in 8860/8861)
Temperature
Measurement Range
General 10°C/division (-100°C/ to +200°C), 50°C/division (-200°C/ to +1000°C)
100°C/division (-200°C/ to +2000°C), 3 ranges, full scale: 20 divisions
digital filter 50/60/10 Hz, measurement resolution 1/1000 of measurement range (using 16-bit A/D conversion; installed in 8860/8861)
Thermocouple range
(JIS C 1602-1995)
(ASTM E-988-96)
K: -200 to 1350°C, J: -200 to 1200°C, E: -200 to 1000°C, T: -200 to 400°C, N: -200 to 1300°C
 R: 0 to 1700°C  S: 0 to 1700°C B: 400 to 1800°C W (WRe5-26): 0 to 2000°C
Thermocouple range:
reference junction compensation
internal/ external (switchable), line fault detection ON/OFF switchable
Data refresh rate 50 ms/all channels (digital filter OFF), 300 ms/all channels (digital filter 50/60 Hz), 1.4 s/all channels (digital filter 10 Hz)
Accuracy Voltage ±0.2% of full scale
 thermocouple
(K, J, E, T, N)
±0.05% of full scale ±1°C
 thermocouple
(R, S, B, W)
±0.05% of full scale ±2°C (400°C or more), ±0.05% of full scale ±3.5°C (less than 400°C)
reference junction compensation accuracy  ±1°C (added to measurement accuracy with internal reference junction compensation)
Max. allowable input 40 V DC (the maximum voltage that can be applied across input pins without damage)
Dimensions and Weight approx. 170 (6.69in) W × 20 (0.79in) H × 183 (7.20in) D mm, approx. 385 g (13.6 oz)
 
Ordering Information
Model Description Buy Online Price
8958 Hioki 8958 16ch Scan Unit TC/DC Volts (8860/01) Buy  Now $1,695.00

Included Accessories

 
  • 1 - Flathead screwdriver
  • 2 - Short bar
 

Compatible Cords

9197 Hioki 9197 Test Lead Buy  Now $109.00
9198 Hioki 9198 Test Lead Buy  Now $69.00
9665 Hioki 9665 Test Probe (10:1) for 8855 Buy  Now $199.00
9666 Hioki 9666 Test Probe (100:1) for 8855 Buy  Now $249.00
9217 Hioki 9217 Connection Cord (BNC-BNC) Buy  Now $89.00

Hioki Catalog   |   Hioki Price List

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved