| Specifications |
|
Specifications |
Description |
|
Measurement Functions |
Number of Channels |
2 |
|
Type of Measurement |
Voltage
Measurements |
|
Input Connectors |
Voltage input/
integrated preamplifier input |
metallic BNC connector (for voltage
input: input impedance 1MΩ, input capacitance 200pF or
less) |
|
Charge input |
miniature connector |
|
Max. rated voltage to earth |
30Vrms or 60V DC (with input
isolated from the unit, the maximum voltage that can be
applied between input channel and chassis and between
input channels without damage) |
|
Suitable transducer |
Charge input |
Charge-output type piezoelectric
acceleration pick-up sensor |
|
Internal preamp input |
Acceleration pick-up sensor with an
internal preamp |
Measurement Range
Charge input (miniature connector) Internal pre-amp input
(BNC connector) |
General |
50m (m/s2)/DIV to 10k (m/s2)/DIV, 12
ranges × 6 types, charge input sensitivity: 0.1 to 10 pC/(m/s2) |
|
integrated pre-amplifier input |
0.1 to 10 mV/(m/s2) |
|
amplitude accuracy |
±2% of full scale |
|
frequency characteristics |
f1 to 50kHz, +1/−3dB, |
|
low-pass filter |
500Hz/5kHz, |
|
pre-amplifier drive power source |
2mA ±20%, +15V ±5% |
|
maximum input charge |
±500pC (high-sensitivity setting, 6
ranges), ±50000pC (low-sensitivity setting, 6 ranges) |
Measurement Range
Charge input Voltage input (BNC connector) |
General |
500μV to 2V/DIV, 12 ranges, |
|
DC amplitude accuracy |
±0.4% of full scale |
|
frequency characteristics |
DC to 400kHz, +1/−3 dB |
|
Low-pass filter |
5Hz/500Hz/5kHz/100kHz |
|
input coupling |
DC, GND, AC |
|
Max. allowable input |
30Vrms or 60V DC |
|
Measurement resolution |
1/80 to 1/32 of measurement range
(depending on measurement sensitivity; installed in
8860/8861) |
|
Maximum sampling rate |
1MS/s (2-channel simultaneous sampling) |
|
Anti-aliasing filter |
Integrated filter for suppressing
aliasing distortion caused by FFT processing (automatic
cutoff frequency setting/OFF) |
|
Additional Notes |
When using Model 8947 with serial number
earlier than 040933650 on Models 8860 or 8861, residual
noise will be 200 μVp-p. |