| Specifications |
|
Specifications |
Description |
|
Measurement Functions |
Number of Channels |
2 |
|
Type of Measurement |
Distortion
Measurements |
|
Temperature measurement with with
Thermocouple |
|
Input Connectors |
Voltage input |
Isolated BNC |
|
Thermocouple input |
Plug-in Terminal |
|
Voltage Measurement Range |
0µV to 2V/division; 12 settings, the
measurement resolution is 1/80 of range
(When attached to recorder where full-scale (f.s.) = 20
divisions) |
1 mV to 5 V/division; 12 settings, the
measurement resolution is 1/160 of range
(When attached to recorder where full-scale (f.s.) = 10
divisions) |
|
Low-pass filter: 5/500/5 k/100 kHz |
|
Temperature Measurement Range |
|
10 °C to 100 °C/division; 4 settings, the
measurement resolution is 1/80 of range
(When attached to recorder where full-scale (f.s.) = 20
divisions) |
20°C to 200°C/division; 4 settings, the
measurement resolution is 1/160 of range
(When attached to recorder where full-scale (f.s.) = 10
divisions) |
|
Low-pass filter |
5/500 Hz |
|
Thermocouple Range |
K: -200 to 1350 °C |
E: -200 to 800 °C |
J: -200 to 1100 °C |
T: -200 to 400 °C |
|
N: -200 to 1300 °C |
R: 0 to 1700 °C |
S: 0 to 1700 °C |
B: 300 to 1800 °C |
|
Maximum sampling rate |
Voltage input: 1 MS/s, Temperature
measurement: 4 kS/s (2-channel simultaneous sampling) |
|
Accuracy |
Voltage input |
DC amplitude ±0.4% of f.s. Zero-position
±0.15% of f.s. |
|
Temperature measurement (K, E, J, T, N) |
±0.1% of f.s. ±1 °C, ±0.1% of f.s.
±2 °C (-200 to 0 °C), (R, S): ±0.1% of f.s. ±3 °C, (B):
±0.1% of f.s. ±4 °C (400 to 1800 °C) |
|
Reference junction compensation accuracy |
±0.1 % f.s. ±1.5 °C (internal
compensation) |
|
Zero-position |
Voltage input |
-50 % to 150 %, 1 % steps
(With zero-adjust function) |
|
Temperature measurement |
-100 % to 100 %,
1 % steps |
|
Frequency characteristics |
Voltage input |
DC to 400 kHz + 1/-3 dB
|
|
Temperature measurement |
DC to 1 kHz +
1/-3 dB |
|
Max. allowable input |
30 Vrms or 60V DC (upper voltage which
when applied to between input pins does not damage them) |
|
Max. rated voltage to earth |
30 Vrms or 60V DC (upper voltage which
when applied to between input pins does not damage them) |
|
Dimensions and Weight |
approx. 170 (6.69in) W × 20 (0.79in) H ×
148.5 (5.85in) D mm
approx. 290 g (10.2 oz) |