| Specifications |
| Rated current |
30 A (requires derating at frequency)
|
| Frequency bandwidth |
DC to 50 MHz (-3 dB)
|
| Rise time |
=<7 ns
|
| Noise level |
2.5 mA max. (band width limited to 20 MHz) |
| Continuous maximum input range |
30 Arms |
| Maximum peak current value |
Non-continuous 50A peak |
| Output Voltage rate |
0.1 V/A |
| Amplitude Accuracy |
±1.0% rdg. ±1 mV (0 to 30 Arms / DC, 45 to 66
Hz)
±2.0% rdg. (30Arms to 50 Apeak / DC,45 to 66 Hz) |
| Noise |
2.5 mArms or less (measured with 20 MHz bandwidth
equipment) |
| Sensitivity Temperature Characteristics |
Within ±2% (At 50 Hz/30 Arms input, 0 to 40°C
[32 to 104°F]) |
| Maximum Rated Power |
5.6 VA (Input within the maximum input range.) |
| Max. circuit voltage |
300 V CAT I (insulated wire)
|
| Temperature & Humidity |
Operating Temp. |
0 to 40°C [32 to 104°F] |
| Operating Humidity |
80% rh or less(no condensation) |
| Storage Temp. |
-10 to 50°C [14 to 122°F] |
| Storage Humidity |
80% rh or less(no condensation) |
| Effect of External Magnetic Fields |
Max. 20 mA (equivalent) (DC and 60 Hz, Magnetic
field of 400 A/m) |
| Max. rated voltage to earth |
300 V, CAT-I (insulated conductor) |
| Measurement conductor |
Diameter max. 5 mm [0.20"] |
| Power supply |
Unit |
±12 V ±0.5 V, 5.3 VA max. |
| Plug |
FFA.0S.304.CNAC42Z / LEMO inc. |
| Dimensions, |
Unit |
175 mm(6.89 in)W × 18 mm(0.71 in)H × 40 mm(1.57
in)D, 230 g (8.1 oz), |
Cord Length,
Sensor Cable |
1.5 m (4.92 ft) |
Cord Length,
Power Cable |
Approx. 1 m [39.37"] |