Test Equipment Depot Top Nav Bar
Test Equipment Depot - 800.517.8431 Free Ground Shipping Offer
<< Hart Scientific Probes

Hart Scientific 5616
NIST-traceable Secondary PRT Sensor Probe

  • Temperature range: –200 °C to 420 °C
  • Excellent stability: ±10mK
  • Reference-grade platinum sensing element
  • NIST-traceable calibration included
  • The 5616-12 is a 100-ohm platinum resistance thermometer (PRT) with excellent short-term repeatability
  • Short-term repeat-ability better than ± 0.010 °C (± 0.004°C is typical)
  • Specified to drift no more than ±0.010°C at the triple point of water when exposed up to its maximum temperature (420°C) for 100 hours
  • Sealed INCONEL® 600 sheath is 298 mm (11.75 in) long and 6.35 mm (0.250 in) in diameter
  • Teflon®-jacketed cable is made of silver plated copper that ends with four-wire leads, which eliminate the effects of lead-wire resistance on measurements
     Hart Scientific 5616 NIST-traceable Secondary PRT Sensor Probe!Click for larger image
Fluke HART 5616 Shown
Product Information
Datasheet 417 KB
User's Guide 181 KB
PDF Help
 
You won’t find another NIST-traceable reference temperature sensor that matches the accuracy and temperature range of the 5616 for the same price.

The 5616-12 is a 100-ohm platinum resistance thermometer (PRT) with excellent short-term repeatability and comes with an unaccredited NIST-traceable calibration.

The temperature range of the 5616 covers –200°C to 420°C, and its high-purity platinum element and durability make it great for calibrating in the lab or in the field. When choosing a reference with a platinum element, there are two things you want to look at carefully: the short-term repeatability and the long-term drift. When PRTs are thermally cycled over their temperature range as they would be during a calibration, their resistance at the triple point of water can move up and down within an expected range. Hart Scientific defines this range (called “short-term repeatability”) as the repeatability at the triple point of water during three ther-mal cycles. 5616s are among the best performing in their class with short-term repeat-ability better than ± 0.010°C (± 0.004°C is typical). In addition, the 5616 is specified to drift no more than ±0.010°C at the triple point of water when exposed up to its maximum temperature (420°C) for 100 hours. These specifications are given at k=3 and therefore include a 99.8 % confidence level.

The 5616’s sealed INCONEL® 600 sheath is 298 mm (11.75 in) long and 6.35 mm (0.250 in) in diameter. The probe’s Teflon®-jacketed cable is made of silver plated copper that ends with four-wire leads, which eliminate the effects of lead-wire resistance on measurements.

Use the 5616 with Hart’s 1560 Black Stack, 1529 Chub-E4, or 1502A Tweener thermometer readouts.

Each sensor comes with a manufacturer’s report of calibration. The report includes the expanded uncertainty (k=2) at seven calibration temperature points, ITS-90 calibration coefficients, and a temperature vs. resistance table presented in 1 °C increments. Compare the 5616 to other Secondary Reference PRTs. You’ll like its price, but you’ll love its performance.

 
Specifications
Temperature Range –200 °C to 420 °C
Nominal Resistance at 0.01 °C 100 Ω ± 0.5 Ω
Temperature Coefficient 0.003925 Ω/Ω/°C nominal
Short-term Repeatability ± 0.010 °C at 0.010 °C
Drift ± 0.010 °C at 0.010 °C
Hysteresis ± 0.010 °C maximum
Sensor Length 50.8 mm (2.0 in)
Sensor Location 6.0 mm ± 2.5 mm from tip (0.24 in ± 0.10 in)
Sheath Diameter Tolerance ± 0.08 mm (± 0.003 in)
Sheath Material INCONEL® 600
Minimum Insulation Resistance 500 MΩ at 23 °C
Transition Junction Temperature Range –50 °C to 150 °C
Transition Junction Dimensions 76.2 mm x 9.5 mm (3.00 in x 0.375 in)
Minimum Immersion Length
(< 5 mK error)
102 mm (4.0 in)
Maximum Immersion Length 254 mm (10 in)
Response Time 8 seconds typical
Self Heating (in 0°C bath) 60 mW/°C
Lead-wire Cable Type Teflon®-jacketed cable, Teflon® insulated conductors, 24 AWG stranded, silver plated copper
Lead-wire Length 182.9 cm ± 2.5 cm (72.0 in ± 1.0 in)
Lead-wire Temperature Range –50 °C to 150 °C
Calibration NIST-traceable calibration
 
Calibration Uncertainty
Temperature Expanded Uncertainty (k=2)
–197 °C 0.012 °C
–80 °C 0.012 °C
–38 °C 0.011 °C
0 °C 0.009 °C
156 °C 0.011 °C
230 °C 0.013 °C
420 °C 0.021 °C
 
Ordering Information
Model Size
Length x Diameter
Type Temp. Range Buy Price
561612B 12 in x 1/4 in Bare Wire –200 to 420 °C Buy Now $940.00
561612D 12 in x 1/4 in 5-pin DIN
for Tweener Thermometers
–200 to 420 °C Buy Now $940.00
561612I 12 in x 1/4 in INFO-CON
for 1521/1522
–200 to 420 °C Buy Now $940.00
561612L 12 in x 1/4 in Miniature Spade Lugs –200 to 420 °C Buy Now $940.00
561612S 12 in x 1/4 in Standard Spade Lugs –200 to 420 °C Buy Now $940.00
Optional Accessories
2601 Fluke Hart 2601 Probe Carrying Case Buy Now $110.00

Hart Scientific Catalog   |  Hart Scientific Price List

$40.00 Minimum on Total Order