HOME > DRUCK > LOOP-CALIBRATORS > UPSIII_IS druck logo

GE Druck UPS-III IS
Intrinsically Safe Current Loop Calibrator

Click here for larger image of the GE Druck UPS-III IS Compact Loop CalibratorClick for larger image
GE Druck UPS-III IS Shown
Product Information
Datasheet 771 KB PDF Datasheet
Click here for Information Regarding PDF'sPDF Help
  • ATEX/IEC intrinsically safe
  • Measure or source 0 to 24 mA
  • Accuracy 0.015% of reading
  • Dual mA and % readout, linear or flow
  • Step, span check, valve check, ramp
  • 50 VDC measurement and continuity
  • HART compatible

The GE Druck UPS-III-IS is an intrinsically safe loop calibrator, with its size making it rugged and extremely compact. The GE Druck UPSIII-IS measures 3.5 in x 5.5 in x 1.7 in (90 mm x 140 mm x 42 mm) and weighs just 16.2 oz (460 g). It is an essential tool for loop testing, instrument maintenance and valve set-up, with an easy-to-read display and simple-to-use time saving features.

Features
Graphic Display The graphic display with menu-driven interface is easier to use than traditional knobs, switches, multifunction keys and dual key sequences.
Measure or Source 0 to 24 mA An internal loop supply is available in both measure and source modes, essential during plant shutdowns.
Accuracy 0.015% of Reading This includes 12 month stability and temperature effects. The UPS III IS can maintain a 4:1 calibration ratio with the latest instrumentation and is typically 10 to 40 times more accurate than a Digital Multi-Meter (DMM).
Simultaneous % Readout Displays mA and the percentage value of 4 to 20 mA or 0 to 20 mA. The UPS III IS also converts mA readings to %flow. This is not possible with many loop calibrators.
Step, Span Check and Value Check
Step Mode 0% 25% 50% 75% 100%
4 to 20 mA linear 4 8 12 16 20
0 to 20 mA linear 0 5 10 15 20
4 to 20 mA flow 4 5 8 13 20
0 to 20 mA flow 0 1.25 5 11.25 20
4 to 20 mA
span check
4 - - - 20
0 to 20 mA
span check
0 - - - 20
4 to 20 mA valve 3.8, 4,
4.2
- 12 - 19, 20,
21
Step quickly through linearity tests and zero/span adjustments. In valve mode, seating can be checked and the positioner set.
Auto Step and Ramp Outputs have adjustable rates (1 to 599s) for single-handed loop testing, valve set-up and slew rate checks.
"Fast Set" Output Allows values to be set to within 0.001 mA using the arrow keys. Holding a key quickly ramps the output to the next value. The system is ideal for alarm trip tests.
50 VDC Measurement For loop diagnostics and maintenance of voltage output instruments.
Hart Compatible Internal 235 Ωloop resistor (menu selectable)
EMC Compliance to EN 61326 Complies with heavy industry standard. Most loop calibrators are only suitable for light industrial use.
Other Features
  • Adjustable display contrast
  • Adjustable resolution
  • Uses industry standard AA batteries
  • Battery voltage and battery low indicator
  • Information screen reports serial number, software version and calibration date
  • Primary reading in mA or %
  • Auto power off (enable or disable)
  • Pin protected, closed case calibration
  • Leather case with carry strap. (Note: the UPS III IS must be used in the case for hazardous area use.)
 
Specifications
Function Range Resolution 1 Year Accuracy
% rdg + counts
Remarks
Source mA 24 mA 0.001 0.015% + 2 V-max 30 V
Source mA and 21V R-max 650Ω at 20 mA
Measure mA V-max 30 V
Measure mA and 21V R measure and 21V 15Ω
Measure V 50 V 0.015% + 4 R measure 1MΩ
Continuity <100Ω - 0.5 mA current
 
Ordering Information
Model Description Buy Online Price
UPS-III-IS GE Druck UPS-III IS Intrinsically Safe Compact Loop Calibrator  Buy Now $1,329.00
Included Accessories
  • 1 - Leather Carrying Case with Carry Case
  • 1 - User's Manual
  • 1 - Calibration Certificate
  • 1 - Set of 1.5V AA Batteries
Included