- Voltage Measurements:
- AC : True RMS, up to 600Vrms @ 40 ~ 1K Hz
- DC : up to 600V
- Input Impedance : 1M-Ohm
- Current Measurements:
- AC : True RMS, up to 2Arms @ 40 ~ 1K Hz
- DC : up to 2A
- Current Shunt : 0.1 Ohm @ >20mA; 10 Ohm @ ≤ 20mA
- Diode/Audible Continuity Checks:
- Open Circuit Voltage: 5Vdc
- Short Circuit Current: 2.5mA
- Beep On: ≤25Ω
- Beep Off: ≥50Ω
- LCR Measurement Test conditions:
- Frequency : 100Hz/ 120Hz/ 1KHz/ 10KHz/ 100KHz/ 200KHz
- Level : 1Vrms/ 0.25Vrms/ 50mVrms/ 1VDC (DCR only)
- LCR Measurements:
- Measurement Parameters:
Z, Ls, Lp, Cs, Cp, DCR, ESR, D, Q and θ
- Basic Accuracy : 0.1%
- Dual Liquid Crystal Display
- Auto Range or Range Hold
- USB Interface Communication
- Open/Short Calibration
|
 
BK Precision 889B Shown
|
| The B&K Precision Corp. 889B
Synthesized In-Circuit LCR/ESR Meter is a high accuracy test
instrument used for measuring inductors, capacitors and resistors
with a basic accuracy of 0.1%. Also, with the built-in functions of
DC/AC Voltage/Current measurements and Diode/Audible Continuity
checks, the 889B can not only help engineers and students to
understand the characteristics of electronics components but also
being an essential tool on any service bench. The 889B is
defaulted to auto ranging. However, it can be set to auto or manual
ranging by pressing the Range Hold key. When LCR measurement mode is
selected, one of the test frequencies, 100 Hz, 120 Hz, 1 KHz, 10
KHz, 100 KHz or 200 KHz, may be selected on all applicable ranges.
One of the test voltages, 50mVrms, 0.25 Vrms, 1 Vrms or 1 VDC (DCR
only), may also be selected on all applicable ranges. The dual
display feature permits simultaneous measurements. When DC/AC
voltage/current measurement mode or the Diode/Audible Continuity
Check mode is selected, only the secondary display will be used to
show the result of the measurement.
The highly versatile 889B can perform virtually all the functions
of most bench type LCR bridges. With a basic accuracy of 0.1%, this
economical LCR meter may be adequately substituted for a more
expensive LCR bridge in many situations. Also, with the basic
accuracy of 0.4% in voltage and current measurements, the 889B
performs the functions of a general purpose Digital Multi-Meter and
can be used to replace the DMM on a service bench.
The 889B has applications in electronic engineering labs,
production facilities, service shops, and schools. It can be used to
check ESR values of capacitors, sort and/or select components,
measure unmarked and unknown components, and measure capacitance,
inductance, or resistance of cables, switches, circuit board foils,
etc. |