Features |
Range |
Accuracy |
| AC
Voltage |
50Hz - 500Hz |
400.0 mV |
4.0% + 4 digits |
|
Selection by RANGE button manually, and is specified
from AC 40mV (AC 60mV for True RMS model ACD-6 TRMS-PRO) and up
|
| 50Hz -- 60Hz
|
4.000V, 40.00V, 400.0V |
1.0% + 4 digits |
| 60Hz -- 500Hz |
4.000V, 40.00V, 400.0V |
1.5% + 4 digits |
| 50Hz -- 500Hz |
600V |
2.0% + 4 digits |
| CMRR |
>60dB @ DC to 60Hz, Rs=1kΩ |
| Input
Impedance |
10MΩ, 30pF nominal |
|
DC Voltage |
General Stats |
400.0 mV
4.000V, 40.00V, 400.0V
600V |
0.3% + 3 digits
0.5% + 3 digits
1.0% + 3 digits |
| NMRR |
>50dB @ 50/60Hz |
| CMRR |
>120dB @ DC, 50/60Hz, Rs=1kΩ |
| Input
Impedance |
10MΩ, 30pF nominal (1000MΩ for 400.0mV range) |
| DC Current
(Clamp-on) |
400.0A
|
0.0A ~ 400.0A |
1.5% + 4 digits |
|
1000A |
400A ~ 800A
800A ~ 900A
900A ~ 1000A |
1.5% + 4 digits
2.0% + 4 digits
5.0% + 30 digits |
|
Accuracy: Induced error from adjacent
current-carrying conductor: < 0.01A/A
Accuracy: ∆ Zero mode must be applied to zero the unit before
measurement to ensure accuracy. This is particularly important below
20A. |
| DC Current
(Clamp-on) |
400.0A
|
15Hz ~ 40Hz
40Hz ~ 200Hz
200Hz ~ 400Hz @ <50A*
400Hz ~ 1kHz @ <50A* |
2.0% + 5 digits
1.5% + 5 digits
1.5% + 5 digits
2.0% + 5 digits |
| 800A
|
15Hz ~ 40Hz @ 400A ~ 600A
40Hz ~ 100Hz @ 400A ~ 600A
15Hz ~ 60Hz @ 600A ~ 800A |
2.0% + 5 digits
1.5% + 4 digits
5.0% + 30 digits |
|
Accuracy: Induced error from adjacent
current-carrying conductor: < 0.01A/A
*: Accuracy is specified at <50A in this frequency bandwidth due to
limited calibrator output capability for testing. |
|
Audible Continuity Tester |
Open Circuit
Voltage |
0.4VDC typical |
| Range:
400.0Ω; Accuracy |
1.5% + 6 digits |
| Audible
threshold |
between 10Ω and 120Ω |
|
Diode
Tester |
Open Circuit < 1.6 VDC |
|
Voltage Test Current (Typical) 0.4mA |
|
Capacitance |
500.0nF, 5.000μF, 50.00μF, |
|
500.0μF, 3000μF |
3.5% + 6d |
Range: Additional 50.00nF range accuracy is not
specified
Accuracy: Accuracies with film capacitor or better
Accuracy: Specified with battery voltage above 2.8V (approximately
half full battery). Accuracy decreases gradually to 12% at low
battery warning voltage of approximately 2.4V |