Infrared Temperature Calibrator, 28 to 400°C (28 to 752°F)

  • Dry Block Calibrator (-9 to 311°F) -23 to 155°C
  • Features improved Peltier elements that use a "Multi-Stage Technology" that improves the efficiency, and extends the life of the heating/cooling block
  • Enhanced accuracy and stability-MVI circuitry ensures temperature stability despite mains supply variations
  • Timesaving features- Fast one-key-one-function access to the automatic switch test and the step function
  • Documentation made easy- RS-232 communication and calibration software AMECAL-TEMPERATURE are standard
  • High accuracy and long term stability-specified drift over a one year period of time

Click here for larger image - Ametek ETC125A Easy Temperature CalibratorClick here for larger image - Ametek ETC125A Easy Temperature Calibrator
Ametek ETC400R Shown

Product Information
Datasheet 1.34 MB
Ordering Info Jump to Product Models, Prices, Accessories, and Ordering Information

PDF Help

Designed for people who perform tests and verifications of temperature sensing devices in the field. This instrument is ideal when time is a critical factor and the highest accuracy is not a critical factor. Reduced size and weight are important considerations because the unit is able to fit into a tool box or instrument carrying case and can be used for sensors that are difficult to access.

One-key-one-function user interface provides immediate access to setting the temperature and the auto-step timesaving function. There is no need for manipulation of sophisticated menus. The Stability indicator provides audible and visual prompts when the temperature is stable. This function also includes a 3 minute countdown before the stable condition.

Stainless steel and rubber side panels make the instrument suitable for many years of faithful duty in an industrial environment.

ETC-400 R for infrared thermometers
The ETC-400 R is designed for optimum speed in connection with calibration of infrared thermometers. The 36 mm target provides the optimum size for reliable calibration of infrared thermometers in the process industry as it is designed for high accuracy and long-term stability while maintaining speed.

With regard to the coating of the target it has been especially designed for space technology applications, which secure long time performance under high temperature influence. In combination with the shape of the target it ensures the emissive rate of 0.96.

Easy-to-use, intuitive operation
All instrument controls are accessed directly from the front panel. The main functions on the ETC series are designed with one-key-one-function logic. This means that there are no difficult multiple keystrokes to remember to access primary functions. The easy-to-read, backlit display features dedicated icons, which help in identifying instrument conditions and operational steps.

Set temperature
The ’’Up’’ and ’’Down’’ arrow keys allow the user to set the exact temperature desired with a resolution of 0.1°C or °F.

Instrument setups
The ETC-series stores the complete instrument setup, including: engineering units, stability criteria, resolution, auto-step settings, and maximum temperature.

Stability indicator
The bold checkmark on the display indicates that the calibrator has reached the desired set temperature and is stable. The operator may change the stability criteria and establish a greater level of confidence in the calibration results as desired. A convenient countdown timer is activated three minutes before the unit reaches stability. This prompts you to be prepared to record results.

Auto-stepping
This feature saves time. The operator may stay in the control room, or another remote location, monitoring the out-put from the sensor-under-test while the ETC- series calibrator is placed in the process and automatically changes the temperature using a programmed step value and rate. Up to 9 different temperature steps may be programmed, including the hold time for each step. This feature is also ideal for burning-in new sensors prior to installation; this minimizes initial drift and allows for initial testing. It is also useful for testing temperature data loggers.

Maximum temperature
From the setup menu, you can select a lower maximum temperature limit for the calibrator. This function prevents damage to the sensor-under-test caused by the application of excessive temperatures.

JOFRA IR-LAB software for the ETC-400 R
As an extra feature the ETC-400 R will be delivered with a small mathematical program, which will constitute a powerful tool together with the calibrator. The program enables you to calculate at which temperatures you need to calibrate, if your IR thermometer is either locked to a fixed emission factor or if you just want to calibrate your thermometer at a certain emission factor. The program facilitates the whole issue of correcting settings of emission factors and temperatures.

The calibration surface of the JOFRA ETC-400 R IR calibrator has an emission factor of 0,96. If your IR-thermometer is using a different emission factor than 0.96, the result will be a faulty temperature reading on your IR thermometer. However if your IR thermometer is using an emission factor of 0.95 or 0.98 – a helpful diagram is part of the standard delivery.

Re-calibration/adjustments made easy
The ETC- series has a very easy and straightforward procedure for re-calibration/adjustment. There is no need for a screwdriver or PC software. The only thing you need is a reliable reference thermometer. Place the probe in the calibrator and follow the instructions on the display.

Calibration of up to 24 sensors with Ametek ASM
Using the Ametek ITC series together with the ASM Advanced Signal Multi-scanner offers a great time-saving automatic solution to calibrate multiple temperature sensors at the same time.

The ASM series is an eight channel scanner controlled by JOFRACAL software on a PC. Up to 3 ASM units can be stacked to calibrate up to 24 sensors at the same time. It can handle signals from 2-, 3- and 4 wire RTD’s, TC’s, transmitters, thermisters, temperature switches and voltage.
For more info on the ADM Advanced Signal Temperature Multi-Scanner, please visit to the 7533PC page.
 
ETC-400R Specifications
Mains specifications
Voltage Multi-voltage 115VAC and 230VAC
115V(90-127)
230V(180-254)
Frequency 45 - 65 Hz
Power consumption (max) 75 VA
Temperature range @ ambient temp. 23°C / 73°F 28 to 400°C / 82 to 752°F@ 23°C
Resolution Selectable 1º or 0.1º
Stability ±0.3°C / ±0.54°F
Time to stability 3 Minutes
Resolution 1° or 0.1°
Accuracy +0.5°C / +0.9°F
110 mm / 4.3 in
Auto stepping Programmable Up to 9 Steps
Dwell time on each step Programmable
Multi-information display Stability indicator Clear Checkmark
Countdown timer before stable 3 Minutes
Temperature SET and READ simultaneously
Alphanumeric messages Yes
Calibration status icons Yes
Service facilities
Adjustment of the unit from the keypad Yes
Self explaining guide in display Yes
Other information Display serial number, software revision level, and last calibration date
Setup facilities
Stability criteria Extra time before "stable indication" is shown
Display resolution 1° or 1°C/°F
Temperature units °C or °F
Slope rate 0.1 to 9.9°/minute
Maximum temperature Any value within range
Temperature range
Range 33 to 650°C / 91 to 1202°F
Heating time 28 to 200°C / 82 to 392°F 2 minutes
200 to 400°C / 392 to 752°F 3 minutes
Cooling time 400 to 200°C / 752 to 392°F 9 minutes
200 to 50°C / 392 to 122°F 24 minutes
Training mode (heating/cooling block disabled) Simulation of all functions Yes
Simulating heating and cooling Approx. 100° per minute
Physical
Temperature Operating 0 to 40°C / 32 to 104°F
Storage -20 to 50°C / -4 to 122°F
Humidity 0 to 90% RH
Serial data interface RS232 (9-pin Male)
Instrument Weight 1.8 kg / 3.9 lb
Size (L x W x H) 172 × 72 × 182 mm / 6.8 × 2.8 × 7.2 in
 
Ordering Information
Model Description Buy Online Price
ETC400R  Ametek ETC400R Easy Temperature Series Infrared Temperature Calibrator, 28 to 400°C (28 to 752°F)  Buy Now Sale $2,379.75
(Reg. $2,505.00)
Included Accessories
 
  • 1 - Traceable calibration certificate - temperature performance
  • 1 - Calibration software AMECAL-LIGHT
  • 1 - User and reference manual
  • 1 - Mains power cable
  • 1 - Shoulder strap
  • 1 - RS232 cable
  • 1 - Carrying case
  • 1 - JOFRA IR-LAB calibration software
  • 1 - Emissivity table

Included

Ametek Catalog   |   Ametek Price List

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved