Test Equipment Depot - 800.517.8431 McAfee Secure sites help keep you safe from identity theft, credit card fraud, spyware, spam, viruses and online scams
<< Ametek Compact Pressure Calibrator Index

Ametek CPC010CINDG
Compact Pressure Calibrator, Vacuum to 0.7 bar (10 psi)
Calibrator Indicator Only

  • Easy-to-read dual line LCD display that indicates pressure and mA (loop signals) and percent of full scale simultaneously.
  • Store peak values (upper and lower).
  • Single button zeroing.
  • Ideal for on-site use: Complete ready-to-use calibration systems in one carrying case or as individual indicators.
  • One hand operation: No need for a flat surface to operate the system
  • High flexibility: The calibration system can easily be disassembled so the indicator can be used separately for pressure monitoring tasks.
  • Transmitter loop: Measure the transmitter loop current up to 24mA (125%) and the pressure at the same time.
  • Easy pressure switch test.
  • Record min. and maximum readings.
  • Easy recalibration of the unit.
  • Temperature compensation: ±0.005% F.S. /°C outside 18 to 28°C (±0.003% F.S. /°F outside 64 to 82°F).
  • Overload maximum pressure +50% Full Scale. 
  • Media compatibility Non-corrosive gasses and liquids.
  • Connection 1/8 in. NPT female - see delivery section for adapters.
  • Min/Max Peak hold capture 50 milliseconds
    mA input 0 to 24 mA accuracy: ±0.015% of reading +2 µA.

Click here for larger image - Ametek CPC010CINDG Compact Pressure Calibrator, Vacuum to 0.7 bar (10 psi)Click here for larger image - Ametek CPC010CINDG Compact Pressure Calibrator, Vacuum to 0.7 bar (10 psi)
Ametek CPC010CINDG
Shown

Product Information
Datasheet 1.88 MB
Ordering Info Jump to Product Models, Prices, Accessories, and Ordering Information

PDF Help

The Ametek CPC series is recognized by its high quality and carefully selected features that save time and make calibration tasks easy to per-form; e.g. dual information display, 17 different pressure units, transmitter mA reading, min/max recording, peak reading, and semi-automatic pres-sure switch test finding the closing, opening and the hysteresis pressure (dead band).

Maintenance of the instrument is made easy. Recalibration of the unit can be performed locally with an accurate pressure reference: No need to send the unit back to the manufacturer for re-calibration. For all systems the pump and the indicator are separate so replacement of the pump does not mean that you have to send the complete system for repair.

The Ametek CPC010CINDG can be delivered as an individual indicator or as a complete calibration system ready for use in a carrying case, including pump, hose, connectors, battery, test leads, user manual, and traceable calibration certificate.

This system includes the Ametek APC calibrator together with one of the following pneumatic hand pumps: T-730, T-740 or T-750, depending on the chosen pressure range. System A is an easy-to-use single hand operated calibration system. No need for a flat surface to operate the system - just hold it in your hand.

The CPC010CINDG comes in an aluminum carrying case with cutouts for fittings, hose, and the complete assembled calibration unit - no time is required to assemble the unit every time you need it. The calibrator (indicator) can at any time be used separately for other pressure test tasks. A special quick connector between the pump and the indicator makes it possible to separate the system in seconds and to swivel the indicator for easy viewing.

The Ametek APC delivery always includes a protective softcase for the indicator. All APC indicators used with the CPC010CINDG are delivered calibrated in both pressure and vacuum.
 
Button Functions

Details

Clear dual line display

Large digits display the pressure, icons show current status and mode, and the smaller digits show device-under-test conditions.
Electrical connections The CPC indicator have inputs for mA readings for pressure switches and transmitters.
ON / OFF Auto shut-off to extend battery life (user programmable)
mA% Switches the mA numeric display between mA and percentage of range.
Switch test Turns on the switch test mode. Hold to clear previous test.
Units Easy access to 17 different pressure engineering units. Press and hold while unit is powering up to change the power saver setting.
Zero Keep the high accuracy: ZERO the instrument before every test.
Serial interface Serial interface to communi-cate with a PC (terminal). Also used if adjustment of the unit is necessary.
Max / Min Reads the highest and lowest pressure value since the last reset the CPC. Monitor a safety valve or look for pressure spikes in the process.
Pressure connections 1/8 in NPT female.
 
Dedicated switch test function
All CPC models have a time saving semi automatic pres-sure switch test. The instrument records the closing and opening pressure for the switch. The CPC can also be used to monitor a certain measurement point in the process and display the minimum and maximum value. This feature can be very helpfull when checking pressure spikes around valves.

JOFRACAL  Calibration Software
The JOFRACAL  calibration software ensures easy calibration of RTD´s, thermocouples, transmitters, thermoswithes, pressure gauges and pressure switches. JOFRACAL  can be used with Ametek DPC-500, APC, CPC and IPI pres-sure calibrators, all Ametek temperature calibrators, as well as Ametek AMC900, ASC300 multi signal calibrator and ASM-800 signal multi scanner.
Requirements

Details

Minimum Requirements:
Hardware
  • Intel Pentium II 1.4 GHz Processor
  • 64MB Ram (128MB recommended)
  • Standard VGA (800x600, 256 colors). 1024x768 Recommended
  • CD-Rom Drive for installation of program
  • 1 or 2 free RS-232 serial ports, depending on configurations
Minimum Requirements:
Software
  • Microsoft Windows 98, Microsoft Windows NT 4.0, Microsoft Windows 2000,
  • Microsoft Windows ME, Microsoft Windows XP, Windows Vista\
  • System fonts: MS Sans Serif, Arial
Calibration Software:   View PDF

JOFRACAL  calibration software may also be used for manual calibrations, as it can be set up to accept manual entry of calibration data together with other liquid baths, ice points or dry-block heat sources. The calibration data collected may be stored on a PC for later recall or analysis. The ITC calibrator stores the calibration procedure and may be taken out to the process site without using a personal computer. Once all calibrations are completed, the data may be uploaded to the JOFRACAL  calibration software for post-processing and printing of certificates. The calibration data collected may be stored on the personal computer for later recall or analysis.

 
Specifications
Mains specifications
Pressure range Bar vacuum to 0.7
Psi vacuum to 10
Ranges Metric/Imperial
Detailed Indicator Pressure Ranges Chart >>
Engineering units (built-in) User selectable: 17 Units psi, bar, mbar, kg/cm2, inH2O @ 4°C, inH2O @20°C, inH2O @60°C, mH2O @4°C, mH2O @20°C, cmH2O@4°C, cmH2O @20°C, ftH2O @4°C, ftH2O @20°C, ftH2O@60°C, inHg @0°C, mmHg @0°C, kPa
Pressure accuracy Accuracy positive pressure ±0.05% of F.S.
Accuracy vacuum ±0.1% F.S
Temperature compensation ±0.005% F.S. /°C outside 18 to 28°C
±0.005% F.S. /°C outside 18 to 28°C
Additional Notes F.S. Full Scale is the numerical value of the positive pressure range. The accuracy includes non-linearity, hysteresis, repeat-ability and temperature effect 18 to 28°C / 64 to 82°F.
All above specifications are specified at 23°±3°C / 73°±6°F unless otherwise noticed.
Media compatibility Non-corrosive gasses and liquids
Input / output Min/Max Peak hold capture 50 milliseconds
mA input 0 to 24 mA accuracy ±0.015% of reading +2 µA
Switch test Supply max. 9 V
Resolution 1° or 0.1°
Accuracy ±0.25°C / ±0.45°F
Radial homogeneity (difference between holes) 0.04°C / 0.07°F
Immersion depth 150 mm / 5.91 in
Well diameter 30 mm / 1.18 in
Switch input Test voltage Maximum 5 VDC
Test current Maximum 2.5 mA
Temperature range
Range 33 to 320°C / 91 to 608°F
Heating time 50 to 320°C / 122 to 608°F 7 minutes
Cooling time 320 to 100°C / 608 to 212°F 30 minutes
320 to 50°C / 608 to 122°F 60 minutes
Miscellaneous
Temperature Operating -10°C to 50°C / 14°F to 122°F
Storage -20°C to 60°C / -4°F to 140°F
CE conformity EN50082-1, 1992 and EN55022, 1994 Class B
Safety CSA C22.2 No. 1010.1, 1992
Physical
Instrument Weight 400 g / 14.1 oz
Size (L x W x H) 188 x 84 x 52 mm / 7.37 x 3.31 x 2.05 in
Display Type Dual information LCD plus icon indicators
Update Rate Twice per second
RS232 communication interface Connector Stereo Jack
Serial 0-5 VDC, 9600 baud, 8 data, no parity, 1 stop
Protocol ASCll command language
Pressure Connection 1/8 in. NPT female
Overload Max. Pressure +50% F.S.
"OL" Activation +20% F.S.
Power Supply Battery 9V Alkaline (x1)
Battery Life Min. 15 Hours
Low battery indicator When remaining power reaches 7V
 
Ordering Information
Model Description Buy Online Price
CPC010CINDG  Ametek CPC010CINDG Compact Pressure Calibrator, Vacuum to 0.7 bar (10 psi), Calibrator Indicator Only  Buy Now Sale $1,182.75
(Reg. $1,245.00)
Included Accessories
 
  • 1 - Traceable calibration certificate (NIST) With pressure and vacuum performance
  • 1 - 9 V Battery
  • 1 - Converter to 1/4 in NPT female
  • 1 - Converter to 1/4 in BSP female
  • 1 - Test leads; red and black
  • 1 - Connection cable for RS232
  • 1 - JOFRACAL calibration software
  • 1 - Soft carrying case
  • 1 - User manual

Included

Ametek Catalog   |   Ametek Price List

$20.00 Minimum on Total Order

Home Page | Test Equipment Catalog | Specials | Repair
We Buy Surplus | Ordering/Contact Info
____________________________________________________________

99 Washington Street, Melrose, MA 02176
781.665.1400 | 800.517.8431 | Fax: 781.665.0780 | Email:

Copyright © Fotronic Corporation All Rights Reserved