Advanced Pressure Calibrator, 0 to 70 bar (1,000 psi),

  • High accuracy - ±0.04% of reading ±0.01% F.S. with no low range degradation in accuracy
  • True field calibrator
    Fully temperature compensated - same high accuracy in the lab and in the field
  • Transmitter scaling Calculation of % error of the device-under test with user-defined scaling
  • Easy pressure switch test - Automatically records opening and closing point values on the display
  • Capture the min and max pressure readings for safety valve applications
  • More than 60 different pressure modules, accuracy up to 0.01% F.S.
  • Documentation made easy - RS232 communication and JOFRACAL calibration software are included in the standard delivery
Click here for larger image - Ametek APC01KGDOXG Advanced Pressure Calibrator, 0 to 70 bar (1,000 psi)Click here for larger image - Ametek APC01KGDOXG Advanced Pressure Calibrator, 0 to 70 bar (1,000 psi)

Product Information
Datasheet 1.97 MB
Ordering Info Jump to Product Models, Prices, Accessories, and Ordering Information
PDF Help
The Jofra APC Calibrators are the latest in the new generation of AMETEK handheld pressure calibrators. This series was designed to meet high accuracy pressure calibration applications and to make your tasks easier. The APC10KGoffers features such as dual information display, 17 different pressure units, transmitter supply, scalable mA input for % error calculation, serial communications, and external pressure module capability. The accuracy of the APC calibrators is deadweight caliber throughout the range and is temperature compensated for operation in controlled or in process applications. This is a truly superior pressure calibrator.

The Ametek APM series of pressure modules extends the application base for the APC calibrators by allowing for calibrations in ranges other than those of the indicator; and at a lower cost than purchasing another APC. The APM modules are also compatible with several other Ametek calibrators.

Calibration of the instrument may be performed locally without returning the APC unit to the manufacturer. Everything needed is an accurate pressure reference. If a factory calibration is require, the pump and indicator are independent and only the indicator needs to be returned.


Clear dual line display Large digits display the pressure, icons show current status and mode, and the smaller digits show device-under-test conditions.
mA/% Save time and effort in calculating the error between the APC and the device-under-test by entering the pressure range corresponding to 4-20mA. This gives immediate read-out of the error in %.You can also read mA or % of 4-20mA in the display.
Electrical connections The APC indicator have inputs for mA readings for pressure switches and transmitters, including 24 VDC supply, and for external Ametek APM pressure modules.
ON / OFF Auto shut-off to extend battery life (user programmable).
Switch Type Perform semi-automatic pressure switch tests. The APC records the “close” and “open” pressure values and calculates the hysteresis (deadband). You change the pressure and the APC does the rest
MAX/MIN Reads the highest and lowest pressure value since you last reset the APC. Monitor a safety valve or look for pressure spikes in the process.
Pressure connections 1/8 in NPT female 15 bar (200 psi) and below.
1/8 in NPT male for 35 bar (500 psi) and above.
Units Easy access to 17 different pressure engineering units.
Serial interface Serial interface to communicate with a PC (terminal). Also used if adjustment of the unit is necessary.
Zero Keep the high accuracy: ZERO the instrument before every test.

JOFRACAL  Calibration Software
The JOFRACAL  calibration software ensures easy calibration of RTD´s, thermocouples, transmitters, thermo switches, pressure gauges and pressure switches. JOFRACAL  can be used with Ametek DPC-500, APC, CPC and IPI pres-sure calibrators, all Ametek temperature calibrators, as well as Ametek AMC900, ASC300 multi signal calibrator and ASM-800 signal multi scanner.


Minimum Requirements:
  • Intel Pentium II 1.4 GHz Processor
  • 64MB Ram (128MB recommended)
  • Standard VGA (800x600, 256 colors). 1024x768 Recommended
  • CD-Rom Drive for installation of program
  • 1 or 2 free RS-232 serial ports, depending on configurations
Minimum Requirements:
  • Microsoft Windows 98, Microsoft Windows NT 4.0, Microsoft Windows 2000,
  • Microsoft Windows ME, Microsoft Windows XP, Windows Vista\
  • System fonts: MS Sans Serif, Arial
Calibration Software:   View PDF

JOFRACAL  calibration software may also be used for manual calibrations, as it can be set up to accept manual entry of calibration data together with other liquid baths, ice points or dry-block heat sources. The calibration data collected may be stored on a PC for later recall or analysis. The ITC calibrator stores the calibration procedure and may be taken out to the process site without using a personal computer. Once all calibrations are completed, the data may be uploaded to the JOFRACAL  calibration software for post-processing and printing of certificates. The calibration data collected may be stored on the personal computer for later recall or analysis.

Ametek APM module
The Ametek APM external pressure modules extend the range of the Ametek APC010CAXXG. There are more than 60 models available with gauge, absolute, differential, and vacuum pressure references and in metric and imperial engineering units. The modules are engineered for in-plant, field, or laboratory use. They are ready-to-use with the Ametek APC and the protocol allows for immediate recognition and use of the module once plugged into the calibrator.

Mains specifications
Pressure range Bar 0 to 70 bar
Psi 1,000 psi
Ranges Metric/Imperial
Detailed Indicator Pressure Ranges Chart >>
Engineering units (built-in) User selectable: 17 Units psi, inH2O@4°C, inH2 O@20°C, inH2 O@60°F, inHg@0°C, ftH2 O@4°C, ftH2O@20°C, ftH2O@60°F, bar, mbar, kPa, kg/cm 2, cmH 2O@4°C, cmH2O@20°C, mH2O@4°C, mH2 O@20°C, mmHg@0°C
Pressure accuracy Pressure ±0.04% rdg + 0.01%F.S.
Full temperature compensation 0 to 50°C / 32 to 122°F
Media compatibility Non-corrosive gasses and liquids
Input / output Min/Max Peak hold capture 50 milliseconds
mA range 0 to 24 mA
mA accuracy  ±0.015% rdg +2 µA
Transmitter supply 24 VDC
Switch test Supply max. 9 V
RS232 communication interface Connector Stereo Jack
Serial 0-5 VDC, 9600 baud, 8 data, no parity, 1 stop
Protocol ASCll command language
Pressure Connection Indicators 200 psi / 15 bar and below 1/8 in. NPT female
Indicators 500 psi / 35 bar and above 1/8” NPT male
Pressure Overload Max. Pressure +50% F.S.
"OL" Activation +20% F.S.
Ordering Information
Model Description Buy Online Price