| HPC600 Specifications |
| Engineering units (built-in) |
User selectable |
15 units |
Pressure accuracy reference temp. (18 to 28°C / 65 to 82°F) |
All pressure ranges |
±0.020% RDG + 0.015% F.S. |
| Vacuum |
±0.025% F.S. |
Pressure accuracy ambient temp. (-10 to 50°C / 14 to 122°F) |
All pressure ranges |
±0.030% RDG + 0.015% F.S. |
| Vacuum |
±0.035% F.S. |
| Display |
LCD |
Graphical (with light) |
| Display resolution |
5 digits |
| Display update |
3 times per second |
Input / output (18 to 28°C / 65 to 82°F) |
mA range |
0 to 24 mA |
| mA accuracy |
±0.015% RDG + 2 μA |
| Transmitter supply |
24 VDC ±10% |
| Voltage measurement |
0 to 30 VDC |
| Voltage accuracy |
±0.015% RDG + 2mV |
| RDT range (ohms) |
84 to 140 Ω |
| RTD accuracy |
±0.015% RDG + 0.02 Ω |
| RDT range (temp) |
-40 to 155ºC / -40 to 311ºF |
| HART® resistor |
250 Ω |
| Peak hold capture |
50 milliseconds |
| Switch test input |
5V (<1mA) |
| Media compatibility |
Nickel plated brass and 316 stainless steel |
| Environmental |
Operating temperature |
-10 to 50°C / 14 to 122°F |
| Storage temperature |
-20 to 60°C / -4 to 140°F |
| Ingress protection rating |
IP56 |
| Pressure connection |
All calibrators |
1/8” BSP/NPT female (combined) |
| Pressure overload |
Overload alarm |
“OL” in display at approx. +20% F.S. |
| Power supply |
Battery |
8 x 1.5 VDC Alkaline |
| Battery type |
AA, LR6, MN1500, AM3 |
| Battery lifetime |
More than 300 pump cycles to 10 bar/150 psi |
| Low battery indicator |
Yes |
| Instrument dimensions (LxWxH) |
Calibrator |
203x102x61 mm / 8x4x2.4 in |
| HPC400 weight (incl. battery) |
920 g / 32.5 oz |
| Calibrator, shipping |
560x350x180 mm / 22x13.8x7 in |
| Calibrator weight, shipping |
3,2 kg / 113 oz |
| Miscellaneous |
Compliance: EN 61326 : 2006 and CISPR 11, Edition 5.0 - 2009 Class “B” |