| HPC500 Specifications |
| Engineering units (built-in) |
User selectable |
15 units |
Pressure accuracy ambient temp. (18 to 28°C / 65 to 82°F) |
Pressure |
±0.025% RDG + 0.01% F.S. |
| Vacuum |
±0.10% F.S. |
Pressure accuracy ambient temp. (0 to 50°C / 32 to 122°F) |
Pressure |
±0.04% RDG + 0.01% F.S. |
| Vacuum |
±0.05% F.S. |
| Display |
LCD |
Graphical (with light) |
| Display resolution |
5 digits |
| Display update |
3 times per second |
| RS232 communication interface |
Connector |
LEMO |
| Serial |
0-5 VDC, 9600 baud, 8 data, no parity, 1 stop |
| Protocol |
ASCII command language |
Input / output (18 to 28°C / 65 to 82°F) |
mA range |
0 to 24 mA |
| mA accuracy |
±0.015% RDG + 2 μA |
| Transmitter supply |
24 VDC ±10% |
| Voltage measurement |
0 to 30 VDC |
| Voltage accuracy |
±0.015% RDG + 2mV |
| HART® resistor |
250 Ω |
| RTD range (ohms) |
84 to 140 Ω |
| RTD accuracy |
±0.015% RDG + 0.02 Ω |
| RTD range (temperature) |
-40 to 105ºC / -40 to 220ºF |
| RTC accuracy @ 0ºC/32ºF (temperature) |
±0.10ºC / 0.18ºF |
| Peak hold capture |
50 milliseconds |
| Switch test input |
5V (<1mA) |
| Media compatibility |
Nickel plated brass and 316 stainless steel |
| Environmental |
Operating temperature |
-10 to 50°C / 14 to 122°F |
| Storage temperature |
-20 to 60°C / -4 to 140°F |
| Ingress protection rating |
IP54 |
| Pressure connection |
All calibrators |
1/8” BSP female |
| Pressure overload |
Overload alarm |
“OL” in display at approx. +20% F.S. |
| Power supply |
Battery |
4 x 1.5 VDC Alkaline |
| Battery type |
AA, LR6, MN1500, AM3 |
| Battery lifetime |
Typically more than 35 hours |
| Low battery indicator |
Yes |
| Instrument dimensions (LxWxH) |
Calibrator |
201x99x50 mm / 8.3x3.9x2.0 in |
| HPC500 weight (incl. battery) |
635 g / 22.4 oz |
| Calibrator, shipping |
243x180x80 mm/ 9.6x7.1x3.1 in |
| Calibrator weight, shipping |
1.1 kg / 39.5 oz |
| Miscellaneous |
Compliance: EN 61326 : 2006 and CISPR 11, Edition 5.0 - 2009 Class “B” |