HOME > AMETEK > HANDHELD-PRESSURE-CALIBRATORS > HPC500002AGXXXXIND ametek logo
 

Ametek Jofra HPC500002AGXXXXIND
Handheld Pressure Calibrator - 0 to 2 bar / 30 psi absolute
Discontinued

  • Single sensor high accuracy pressure calibrator
  • mA measurement
  • Voltage measurement
  • 24 VDC transmitter supply
  • High speed pressure switch test
  • High speed min./max. hold
  • Temperature measurement with external RTD sensor
  • JOFRA APM pressure module connection
  • RS232 communication
  • On-line calculation of sensor error %
  • Automatic leak test timer function
  • Delta/sum calculation on pressure measurements
  • Display contrast adjustment
  • Instrument setup lock
  • Storage of 5 setups
  • Setup of automatic off timer
  • Setup of no. of display windows/measurement channels
  • Select display high or low resolution
  • Setup of backlight timer
  • HART resistor on/off
  • Normal or dampened display update rate
    Includes:
  • HPC500, 0 to 2 bar / 30 psi absolute
  • NIST traceable calibration certificate

Click here for larger image - Ametek HPC500 Handheld Pressure CallibratorClick here for larger image - Ametek HPC500 Handheld Pressure Callibrator
Ametek HPC500 Shown

Product Information
Datasheet 230 KBpdf
Ordering Info Jump to Product Models, Prices, Accessories, and Ordering Information
pdfPDF Help
HPC500 Specifications
Engineering units (built-in) User selectable 15 units
Pressure accuracy ambient temp.
(18 to 28°C / 65 to 82°F)
Pressure ±0.025% RDG + 0.01% F.S.
Vacuum ±0.10% F.S.
Pressure accuracy ambient temp.
(0 to 50°C / 32 to 122°F)
Pressure ±0.04% RDG + 0.01% F.S.
Vacuum ±0.05% F.S.
Display LCD Graphical (with light)
Display resolution 5 digits
Display update 3 times per second
RS232 communication interface Connector LEMO
Serial 0-5 VDC, 9600 baud, 8 data, no parity, 1 stop
Protocol ASCII command language
Input / output
(18 to 28°C / 65 to 82°F)
mA range 0 to 24 mA
mA accuracy ±0.015% RDG + 2 μA
Transmitter supply 24 VDC ±10%
Voltage measurement 0 to 30 VDC
Voltage accuracy ±0.015% RDG + 2mV
HART® resistor 250 Ω
RTD range (ohms) 84 to 140 Ω
RTD accuracy ±0.015% RDG + 0.02 Ω
RTD range (temperature) -40 to 105ºC / -40 to 220ºF
RTC accuracy @ 0ºC/32ºF (temperature) ±0.10ºC / 0.18ºF
Peak hold capture 50 milliseconds
Switch test input 5V (<1mA)
Media compatibility Nickel plated brass and 316 stainless steel
Environmental Operating temperature -10 to 50°C / 14 to 122°F
Storage temperature -20 to 60°C / -4 to 140°F
Ingress protection rating IP54
Pressure connection All calibrators 1/8” BSP female
Pressure overload Overload alarm “OL” in display at approx. +20% F.S.
Power supply Battery 4 x 1.5 VDC Alkaline
Battery type AA, LR6, MN1500, AM3
Battery lifetime Typically more than 35 hours
Low battery indicator Yes
Instrument dimensions (LxWxH) Calibrator 201x99x50 mm / 8.3x3.9x2.0 in
HPC500 weight (incl. battery) 635 g / 22.4 oz
Calibrator, shipping 243x180x80 mm/ 9.6x7.1x3.1 in
Calibrator weight, shipping 1.1 kg / 39.5 oz
Miscellaneous Compliance: EN 61326 : 2006 and CISPR 11, Edition 5.0 - 2009 Class “B”
 
Ordering Information
Model Description Buy Online Price
HPC500002AGXXXXIND Ametek Jofra HPC500002AGXXXXIND Handheld Pressure Calibrator 0 to 2 bar / 30 psi absolute  Discontinued
Included Accessories
 
  • 1 HPC500 calibrator
  • 1 Traceable calibration certificate (NIST) with pressure and vacuum performance
  • 1 4 x 1.5 volt batteries
  • 1 Adapter to 1/4” NPT male from 1/8” BSP male
  • 1 Adapter 1/4” BSP female from 1/8” BSP male
  • 1 1/8” sealed gasket / washer type
  • 1 Test leads: red and black
  • 1 Hand strap with clip
  • 1 User manual
Included