| Specifications U8002A |
| Electrical Performance |
| Output Ratings (0 °C to 40 °C) |
0 to +30 V, 0 to 5 A |
| Line and Load Regulation |
CV: <0.01% +2 mV, CC: <0.02% +2 mA |
| Ripple and Noise (25 °C ±5 °C) |
CV: 12 mVp-p, <1 mVrms; CC: 3 mArms |
Load Transient Response Time (within 15 mV from full load to half load and from half load to full load) |
<50 μs |
| Programming Accuracy (25 °C ±5 °C) |
<0.35% +20 mV, <0.35% +20 mA |
| Readback Accuracy (25 °C ±5 ° C) |
<0.35% +20 mV, <0.35% +20 mA |
| Meter Resolution |
Voltage: 10 mV, Current: 10 mA |
| Maximum Output Float Voltage |
±240 Vdc |
| Protection Features |
Over Voltage Protection Accuracy ± (% of output + offset) |
<0.5% +0.5 V |
| Over Voltage Protection Programmable Range |
1 V to 33 V |
Over Current Protection Accuracy ± (% of output + offset) |
<0.5% +0.5 A |
| Over Current Protection Programmable Range |
1 A to 5.5 A |
| AC Power Input Specifications |
| Input Power Option (selectable) |
100 Vac ± 10%, 47 to 63 Hz 115 Vac ± 10%, 47 to 63 Hz 230 Vac ± 10%, 47 to 63 Hz |
| Maximum Input Power |
500 VA |
| Fuse |
External, customer assessable |
| Physical Specifications |
| Dimensions (H x W x L) |
88.1 mm x 212.3 mm x 394.3 mm |
| Weight |
8.3 kg |